Table 7. Mini-HDMI Breakout Cable to 6 BNC Signal Descriptions
Signal
Connector Type
Description
CLK IN
BNC female
Used to import an external reference clock
CLK OUT
Used to export the reference clock
PFI 0
Bidirectional PFI line
PFI 1
Bidirectional PFI line
PFI 2
Bidirectional PFI line
PFI 3
Bidirectional PFI line
Configuring the PXIe-5163 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and how they are
configured. MAX is automatically installed with NI-SCOPE.
1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware.
Installed modules appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note
If you do not see your module listed, press <F5> to refresh the list of
installed modules. If the module is still not listed, power off the system, ensure
the module is correctly installed, and restart.
4.
Record the identifier MAX assigns to the hardware. Use this identifier when
programming the PXIe-5163.
5.
Self-test the hardware by selecting the item in the configuration tree and clicking
Self-
Test
in the MAX toolbar.
The MAX self-test performs a basic verification of hardware resources.
PXIe-5163 Self-Calibration
For the PXIe-5163, self-calibration yields the following benefits:
•
Characterizes onboard sample clock phase adjustment circuitry
•
Characterizes onboard intermodule synchronization measurement circuitry
•
Corrects for DC gain and offset errors within the oscilloscope for all ranges and both
input impedance paths based on comparison to a precision, high-stability internal voltage
reference
12
|
ni.com
|
PXIe-5163 Getting Started Guide