NI PXIe-4610 Calibration Procedure
|
© National Instruments Corporation
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification procedures. If the
recommended equipment is not available, select a substitute using the requirements listed in
Table 1.
Table 1.
Recommended Equipment
Equipment
Recommended
Models
Where
Used
Requirements
DMM
NI PXI-4070
Gain and
Residual
Offset
Verification
If this instrument is unavailable, use a
digitizer capable of measuring ±10 V
pk
sine
waves at 1 kHz with an amplitude uncertainty
of 0.1 dB or less, a linearity error of 100 ppm
or less and an offset measurement uncertainty
of 100
μ
V or less.
Function
Generator
NI PXI-4461
Gain and
Residual
Offset
Verification
If this instrument is unavailable, use a
function generator capable of sourcing
±1.0 V
pk
sine waves at 1 kHz with an
amplitude uncertainty of 0.1 dB or less and a
frequency uncertainty of 1% or less.
PXI Express
Chassis
NI PXIe-1062Q
—
—