PXIe-4468 Calibration Procedure
30
ni.com/services
Figure 8.
Initial Connection for AO Common-Mode Gain and Offset-mXLR
1. DMM
4. BNC – mXLR Cable
2. Banana – BNC Adapter
5. PXIe-4468 with mXLR
3 Shorting Cable (Banana – Banana)
Verification Procedure
Complete the following procedure to verify common-mode gain and offset.
Repeat 2 times, once for each channel.
10.
Connect the AO channel under test to the DMM as shown in
Repeat 6 times, once for each test point.
11.
Configure the DMM using the information in
Table 24:
DMM Configuration
Configuration
Value
Function
DC Volts (DCV)
Range
Fixed - 10 V
Number of Power Line Cycles (NPLC)
100