Test Conditions

The following setup and environmental conditions are required to ensure the device meets 
published specifications:

Keep connections to the module as short as possible. Long cables and wires act as antennae, 
picking up extra noise that can affect measurements.

Use shielded copper wire for all cable connections to the module. Use twisted-pair wire to 
eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the measurement circuitry of the 
module is at a stable operating temperature.

Ensure that the PXI Express chassis fan speed is set to HIGH, that the fan filters are clean, 
and that the empty slots contain filler panels. For more information, refer to the 

Maintain 

Forced-Air Cooling Note to Users

 document available at 

ni.com/manuals

.

Calibration Procedure

This section provides instructions for verifying the performance of the NI PXIe-4357. 
The calibration process consists of the following steps:
1.

Initial Setup

—Install the module and configure it in Measurement & Automation Explorer 

(MAX).

2.

Verification

—Verify the existing operation of the module. This step confirms whether the 

module is operating within the published specifications prior to adjustment.

3.

Adjustment Procedure

If necessary, perform an external calibration that adjusts the 

module calibration constants with respect to a known calibration source.

4.

Reverification

—Repeat the verification procedure to ensure that the module is operating 

within the published specifications after adjustment.

Initial Setup

You must configure the module in Measurement & Automation Explorer (MAX) to 
communicate with NI-DAQmx.

Complete the following steps to configure a module in MAX. Refer to the 

NI PXIe-4357 and 

TB-4357 Installation Guide and Terminal Block Specifications

 for complete installation 

instructions.
1.

Install the NI-DAQmx driver software.

Caution

Always have the PXI Express chassis turned off when inserting a module.

2.

Insert the module into an available slot in the PXI Express chassis.

3.

Install the terminal block.

4.

Power on the chassis.

Содержание PXIe-4357

Страница 1: ...cedure 7 Specifications 9 Where to Go for Support 10 Software Requirements Install NI DAQmx 9 5 or later on the calibration computer NI DAQmx includes high level function calls to simplify the task of...

Страница 2: ...use the NI DAQmx driver The NI DAQmx C Reference Help includes information about the functions in the driver Calibration Interval National Instruments recommends a calibration interval of one year fo...

Страница 3: ...mance of the NI PXIe 4357 The calibration process consists of the following steps 1 Initial Setup Install the module and configure it in Measurement Automation Explorer MAX 2 Verification Verify the e...

Страница 4: ...on describing the required module connections Each channel consists of a group of four pins specific to that channel Before connecting or disconnecting the calibrator from the module always set the ca...

Страница 5: ...the following connections to the TB 4357 for the channel you want to verify Refer to Figure 2 for visual reference Refer to the Connecting the Calibrator to the Module section for the pin assignments...

Страница 6: ...cation 15 Compare the measured result for each value to the Upper Limit and Lower Limit values in Table 4 If the results are all between these values the module passes the test 16 Set the calibrator t...

Страница 7: ...etermine the gain and offset adjustment calibration constants Complete the following steps to test the accuracy of the module 1 Set the calibrator to standby mode STBY 2 Make the following connections...

Страница 8: ...ration VI The default password is NI 8 Call the DAQmx Adjust 4357 Calibration VI 9 Set the calibrator to output a resistance of 350 This is the second calibration point 10 Call the DAQmx Adjust 4357 C...

Страница 9: ...module you must reverify the accuracy of the module To do this repeat the Verification section Specifications The specifications in Table 5 are provided for reference only refer to the NI PXIe 4357 S...

Страница 10: ...Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Suppo...