Documentation Requirements
The following documents are your primary references for writing your calibration utility with
NI-DAQmx. You can download the latest version of these documents from the NI Web site at
ni.com/manuals
.
•
The
NI PXIe-4357 and TB-4357 Installation Guide and Terminal Block Specifications
provides instructions for installing and configuring the NI PXIe-4357 module and TB-4357
accessory. This document also contains TB-4357 specifications.
•
The
NI PXIe-4357 User Manual
describes how to use the NI PXIe-4357.
•
The
NI PXIe-4357 Specifications
lists the specifications for the NI PXIe-4357.
•
The
NI-DAQmx Help
includes information about creating applications that use the
NI-DAQmx driver.
•
The
NI-DAQmx C Reference Help
includes information about the functions in the driver.
Calibration Interval
National Instruments recommends a calibration interval of one year for the NI PXIe-4357. You
should adjust the recommended calibration interval based on the measurement accuracy
demands of your application.
Test Equipment
National Instruments recommends that you use the equipment in Table 1 for calibrating an
NI PXIe-4357 module.
Table 1.
Recommended Equipment
Equipment
Recommended Model(s)
Requirements
Calibrator
Fluke 5520A
If this instrument is unavailable,
use a calibrator that can provide
resistance values in the range of
0
Ω
to 400
Ω
with 0.01
Ω
resolution, an accuracy of 90 ppm
or better, and automatic lead-wire
compensation.
PXI Express chassis
NI PXIe-1062Q
—
Connection accessory
TB-4357 terminal block
Screw terminal accessory for the
NI PXIe-4357.