NI PXIe-4339/TB-4339 Installation Guide and Terminal Block Specifications
|
© National Instruments
|
15
More Information
After you install NI-DAQmx, the NI-DAQmx documentation is available from
Start»All
Programs»National Instruments»NI-DAQ
. Additional resources are online at
ni.com/
gettingstarted
.
You can access online device documentation by right-clicking your module in MAX and
selecting
Help»Online Device Documentation
. A browser window opens to
ni.com/
manuals
with the results of a search for relevant documents. If you do not have Web access,
documents for supported modules are included on the NI-DAQmx disc.
Troubleshooting
•
Go to
ni.com/support/install
or
ni.com/kb
.
•
If you need to return your National Instruments hardware for repair or device calibration,
go to
ni.com/info
and enter
rdsenn
to start the Return Merchandise Authorization
(RMA) process.
Specifications (TB-4339/B/C)
Note
NI PXIe-4339 module specifications are located in the
NI PXIe-4339
Specifications
document.
All performance specifications are
typical
unless otherwise noted. These specifications are valid
within the full operating temperature range.
Calibration Interval
TB-4339/B/C .................................................... No calibration
Electrical
Resistors
TB-4339
Quarter bridge completion........... 120
Ω
±0.1%, ±10 ppm/°C
Shunt calibration resistor ............. 50 k
Ω
±0.1%, ±10 ppm/°C
TB-4339B
Quarter bridge completion........... 350
Ω
±0.1%, ±10 ppm/°C
Shunt calibration resistor ............. 100 k
Ω
±0.1%, ±10 ppm/°C
TB-4339C
Quarter bridge completion........... 1 k
Ω
±0.1%, ±10 ppm/°C
Shunt calibration resistor ............. 100 k
Ω
±0.1%, ±10 ppm/°C
Trace resistance (screw terminal to connector)
EX+, EX- .......................................................... <20 m
Ω