NI PXIe-4339 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Table 1.
Recommended Equipment
Equipment
Recommended Model
Requirements
Calibrator
Fluke 5700
If this instrument is unavailable, use a
calibrator that can provide DC voltage
values in the range of 0 to ±10 V with an
accuracy of 50 ppm of output or better
and a DC offset error of 4 µV or better.
PXI Express
Chassis
NI PXIe
-
1062Q
—
Connection
Accessory
TB
-
4339/B/C
—