background image

NI PXIe-4322

 

Calibration Procedure

|

© National Instruments

|

5

Each channel consists of two terminal connections specific to that channel as shown in Table 
and must be verified and adjusted individually. Do not make serial or parallel connections for 
verification or adjustment.

Test Conditions

The following setup and environmental conditions are required to ensure the NI PXIe-4322 
meets published specifications.

Keep connections to the NI PXIe-4322 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-4322 are secure.

Use shielded copper wire for all cable connections to the TB-4322. Use twisted-pairs wire 
to eliminate noise.

Maintain an ambient temperature of 23 °C ± 5 °C.

The NI PXIe-4322 temperature will be greater than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the NI PXIe-4322 measurement 
circuitry is at a stable operating temperature.

Ensure that the PXI Express chassis fan speed is set to HIGH, that the fan filters are clean, 
and that the empty slots contain filler panels. For more information, refer to the 

Maintain 

Forced-Air Cooling Note to Users 

document available at 

ni.com/manuals

.

Initial Setup

Refer to the 

NI PXIe-4322 and TB-4322 Installation Guide and Terminal Block Specifications

 

for information about how to install the software and hardware and how to configure the device 
in Measurement & Automation Explorer (MAX).

Note

When a device is configured with MAX, it is assigned a device identifier. 

Each function call uses this identifier to determine which DAQ device to verify or, 
verify and adjust.

Table 2.  

TB-4322 Analog Signal Names

Signal Name

Signal Description

AO+

Positive output terminal

AO-

Negative output terminal

Содержание PXIe-4322

Страница 1: ...on 6 Voltage Output Mode Accuracy Verification 6 Current Output Mode Accuracy Verification 8 Adjustment 9 Voltage Output Mode Adjustment Procedure 9 Current Output Mode Adjustment Procedure 11 EEPROM...

Страница 2: ...Specifications Contains NI PXIe 4322 specifications and calibration interval NI DAQmx Readme Contains Operating system and application software support in NI DAQmx NI DAQmx Help Contains Information...

Страница 3: ...erminal block provides connections to the NI PXIe 4322 Figure 1 shows the connector location for each TB 4322 channel Table 1 Recommended Equipment Equipment Recommended Model Requirements DMM NI PXI...

Страница 4: ...ach channel or For current verification and adjustment connect the DMM to the TB 4322 as shown in Figure 3 for each channel Figure 2 DMM to TB 4322 Voltage Mode Connections Figure 3 DMM to TB 4322 Cur...

Страница 5: ...be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 15 minutes to ensure that the NI PXIe 4322 measurement circuitry is at a stable operating tempe...

Страница 6: ...of the channels depending upon your desired test coverage Voltage Output Mode Accuracy Verification Complete the following procedure to verify the voltage mode accuracy of the NI PXIe 4322 1 Connect t...

Страница 7: ...sured value to the Upper Limit and Lower Limit values in Table 5 4 Repeat step 3 for all voltage values listed in Table 5 5 Disconnect the DMM from the terminal block 6 Repeat steps 3 through 5 for ev...

Страница 8: ...eate a task b Create and configure the AO channel according to the values in Table 7 Table 6 DMM Current Setup Configuration Value Measurement Current Range 100 A for 0 mA test point Minimum range to...

Страница 9: ...ce of operation required to adjust the NI PXIe 4322 Voltage Output Mode Adjustment Procedure Complete the following procedure to adjust the voltage mode accuracy of the NI PXIe 4322 1 Connect the DMM...

Страница 10: ...mx Adjust SC Express Calibration function and configure it as follows g For each adjustment point from first to tenth repeat steps d to f h For eleventh point change the DMM input range to 1 V then re...

Страница 11: ...to the TB 4322 as shown in Figure 3 for each channel c Call the 4322 instance of the DAQmx Get SC Express Calibration Adjustment Point function to obtain 11 points for adjustment d Call the DAQmx Setu...

Страница 12: ...memory EEPROM is immediately updated If you do not want to perform an adjustment you can update the calibration date and onboard calibration temperature without making any adjustments by initializing...

Страница 13: ...compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Support The National Instruments Web site is your complete resource for technical support At...

Отзывы: