NI PXIe-4322

 

Calibration Procedure

|

© National Instruments

|

3

Test Equipment

Table lists the equipment recommended for the performance verification and adjustment 
procedures. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Connecting the TB-4322

Caution

Always have the NI PXI Express chassis powered off when inserting a 

module into the chassis.

The TB-4322 terminal block provides connections to the NI PXIe-4322. Figure 1 shows the 
connector location for each TB-4322 channel.

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Requirements

DMM

NI PXI-4071

Use a DMM that can provide both voltage and 
current measurements.

Accuracy of 20 ppm or better in the smallest range 
to measure 16 V, and offset error of 2.4 µV or 
better in the smallest range to measure 1 V.

Accuracy of 165 ppm or better in the smallest 
range to measure 20 mA, and offset error of 3 nA 
or better in the smallest range to measure 100 µA.

PXI Express 
Chassis

NI PXIe-1062Q

Connection 
Accessory

TB-4322

Содержание PXIe-4322

Страница 1: ...on 6 Voltage Output Mode Accuracy Verification 6 Current Output Mode Accuracy Verification 8 Adjustment 9 Voltage Output Mode Adjustment Procedure 9 Current Output Mode Adjustment Procedure 11 EEPROM...

Страница 2: ...Specifications Contains NI PXIe 4322 specifications and calibration interval NI DAQmx Readme Contains Operating system and application software support in NI DAQmx NI DAQmx Help Contains Information...

Страница 3: ...erminal block provides connections to the NI PXIe 4322 Figure 1 shows the connector location for each TB 4322 channel Table 1 Recommended Equipment Equipment Recommended Model Requirements DMM NI PXI...

Страница 4: ...ach channel or For current verification and adjustment connect the DMM to the TB 4322 as shown in Figure 3 for each channel Figure 2 DMM to TB 4322 Voltage Mode Connections Figure 3 DMM to TB 4322 Cur...

Страница 5: ...be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 15 minutes to ensure that the NI PXIe 4322 measurement circuitry is at a stable operating tempe...

Страница 6: ...of the channels depending upon your desired test coverage Voltage Output Mode Accuracy Verification Complete the following procedure to verify the voltage mode accuracy of the NI PXIe 4322 1 Connect t...

Страница 7: ...sured value to the Upper Limit and Lower Limit values in Table 5 4 Repeat step 3 for all voltage values listed in Table 5 5 Disconnect the DMM from the terminal block 6 Repeat steps 3 through 5 for ev...

Страница 8: ...eate a task b Create and configure the AO channel according to the values in Table 7 Table 6 DMM Current Setup Configuration Value Measurement Current Range 100 A for 0 mA test point Minimum range to...

Страница 9: ...ce of operation required to adjust the NI PXIe 4322 Voltage Output Mode Adjustment Procedure Complete the following procedure to adjust the voltage mode accuracy of the NI PXIe 4322 1 Connect the DMM...

Страница 10: ...mx Adjust SC Express Calibration function and configure it as follows g For each adjustment point from first to tenth repeat steps d to f h For eleventh point change the DMM input range to 1 V then re...

Страница 11: ...to the TB 4322 as shown in Figure 3 for each channel c Call the 4322 instance of the DAQmx Get SC Express Calibration Adjustment Point function to obtain 11 points for adjustment d Call the DAQmx Setu...

Страница 12: ...memory EEPROM is immediately updated If you do not want to perform an adjustment you can update the calibration date and onboard calibration temperature without making any adjustments by initializing...

Страница 13: ...compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Support The National Instruments Web site is your complete resource for technical support At...

Отзывы: