PXIe-4310 Calibration Procedure

|

© National Instruments

|

3

PXIe-4310 Verification and Adjustment

This section provides information for verifying and adjusting the PXIe-4310.

Test Equipment

Table lists the equipment recommended for the performance verification procedures of the 
PXIe-4310. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Table 1.  

Recommended Equipment for PXIe-4310 Verification

Equipment

Recommended Model

Requirements

Calibrator

Fluke 5700A

If this instrument is unavailable, use a 
high-precision voltage source with an 
accuracy of at least 15 ppm for 
adjustment and verification. The 
source needs an output range of at 
least ± 10 V and an output impedance 
of 

50 

.

PXI Express Chassis

PXIe-1062Q

If this chassis is unavailable, use 
another PXI Express chassis, such as 
the PXIe-1082 or PXIe-1078.

Connection Accessory

TB-4310 (10V)

Содержание PXIe-4310

Страница 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Страница 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Страница 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Страница 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Страница 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Страница 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Страница 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Страница 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Страница 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Страница 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: