PXIe-4310 Calibration Procedure
|
© National Instruments
|
3
PXIe-4310 Verification and Adjustment
This section provides information for verifying and adjusting the PXIe-4310.
Test Equipment
Table 1 lists the equipment recommended for the performance verification procedures of the
PXIe-4310. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Table 1.
Recommended Equipment for PXIe-4310 Verification
Equipment
Recommended Model
Requirements
Calibrator
Fluke 5700A
If this instrument is unavailable, use a
high-precision voltage source with an
accuracy of at least 15 ppm for
adjustment and verification. The
source needs an output range of at
least ± 10 V and an output impedance
of
50
.
PXI Express Chassis
PXIe-1062Q
If this chassis is unavailable, use
another PXI Express chassis, such as
the PXIe-1082 or PXIe-1078.
Connection Accessory
TB-4310 (10V)
—