Chapter 2
Analog Input
©
National Instruments Corporation
2-5
Note
You can calibrate NI 6011E and NI 6030E/6031E/6032E/6033E circuitry for either
unipolar or bipolar polarity. If you mix unipolar and bipolar channels in the scan list and
you are using NI-DAQ, NI-DAQ loads the calibration constants appropriate to the polarity
for which AI channel 0 is configured.
Analog Input Terminal Configuration
To be flexible enough to interface with various signal sources, E Series
devices have three different terminal configurations, also referred to as
input modes: Non-Referenced Single-Ended (NRSE) input, Referenced
Single-Ended (RSE) input, and differential (DIFF) input. Table 2-4
describes the three input configurations.
Refer to the
Connecting Analog Input Signals
section for more information
about using these input configurations.
Table 2-3.
Input Ranges for NI 6023E/6024E/6025E and NI 6034E/6035E/6036E
Input Range
Gain
Resolution
NI 6023E/6024E/6025E
NI 6034E/6035E/6036E
–10 to +10 V
0.5
4.88 mV
305
µ
V
–5 to +5 V
1
2.44 mV
153
µ
V
–500 to +500 mV
10
244
µ
V
15.3
µ
V
–50 to +50 mV
100
24.4
µ
V
1.53
µ
V
Table 2-4.
Analog Input Terminal Configuration
AI Terminal Configuration
Description
DIFF
A channel configured in DIFF mode uses two AI lines.
One line connects to the positive input of the device
programmable gain instrumentation amplifier (PGIA), and the
other connects to the negative input of the PGIA.
RSE
A channel configured in RSE mode uses one AI line, which
connects to the positive input of the PGIA. The negative input
of the PGIA is internally tied to AI ground (AI GND).
NRSE
A channel configured in NRSE mode uses one AI line, which
connects to the positive input of the PGIA. The negative input
of the PGIA connects to the AI sense (AI SENSE) input.
Содержание PXI-6031E
Страница 1: ...PXI 6031E...
Страница 2: ...DAQ E Series E Series User Manual E Series User Manual February 2007 370503K 01...