20
|
ni.com
|
NI PXI-5650/5651/5652 Calibration Procedure
12. Repeat steps 5 through 10 for the remaining frequency points shown in Table 10.
Table 10.
Harmonics at
-50 dBm to 0 dBm Output Power
Device
Frequency
Test Limit (dBc)
PXI-5650/5651/5652
10 MHz
<-15
20 MHz
50 MHz
100 MHz
150 MHz
200 MHz
400 MHz
800 MHz
1.0 GHz
PXI-5651/5652
1.6 GHz
<-20
2.0 GHz
2.5 GHz
3.0 GHz
3.3 GHz
PXI-5652
3.5 GHz
<-20
3.7 GHz
4.0 GHz
4.5 GHz
5.0 GHz
5.5 GHz
6.0 GHz
6.6 GHz
The limits in this table are
typical
values. Typical values describe useful product performance beyond
specifications that are not covered by warranty and do not include guardbands for measurement
uncertainty or drift.