
NI PXI-4110 Calibration Procedure
|
© National Instruments
|
11
To ensure the system has had adequate time to settle, wait three seconds after requesting a new
current before taking a measurement with the DMM. The current programming test limits in
Table 7 are derived from the current programming accuracy specifications.
Table 7.
NI 4110 Output Parameters and Test Limits for Current
Programming Accuracy Verification
Channel(s)
Output
(V)
Current
Range
Iteration
Equivalent
Resistance
(
Ω
)
Output
(A)
Test Limit
(A)
0
6
1 A
1
150
0.020000
±0.004030
2
12
0.250000
±0.004375
3
6
0.500000
±0.004750
4
4
0.750000
±0.006063
5
3
1.000000
±0.008000
1
16
1 A
1
500
0.020000
±0.004030
2
40
0.250000
±0.004375
3
20
0.500000
±0.004750
4
13
0.750000
±0.006063
5
9
1.000000
±0.008000
1
16
20 mA
1
†
25,000
0.000400
±0.000060
2
2,000
0.005000
±0.000068
3
1,000
0.010000
±0.000075
4
670
0.015000
±0.000083
5
500
0.020000
±0.000090
2
-16
1 A
1
500
0.020000
±0.004030
2
40
0.250000
±0.004375
3
20
0.500000
±0.004750
4
13
0.750000
±0.006063
5
9
1.000000
±0.008000