National Instruments Phase Matrix PXI-1440 Скачать руководство пользователя страница 33

Quick Start Guide

Phase Matrix PXI-1440 IF Downconverter Module

 

33

 

Equipment Setup to Confirm Basic Module Operation

Figure 16

Equipment Setup to Confirm Basic Module Operation

IF OUT, IF AUX OUT, IF Center Frequency, and 
Power Check 

1

Connect the equipment as shown in 

Figure 16

.

Connect the RF Output of the Signal Generator

to IF A IN of the PXI-1440 IF Downconverter Module.

Connect the REF A OUT of the

PXI-1450B Local Oscillator Module

to REF A IN of the PXI-1440 IF Downconverter Module.

Connect the IF AUX OUT of the

PXI-1440 IF Downconverter Module

to the RF Input of the Spectrum Analyzer.

2

Set the Signal Generator frequency to 249.4 MHz and the 
output power to -2 dBm.

3

Confirm that the output frequency on the Spectrum Analyzer is 
21.4 MHz and the power level is -2 dBm +/- 2 dB.

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

PXI-1440

LOCAL

OSCILLATOR

The rain in Spain

Falls mainly on the plains

TO IF A IN

TO REF A IN

FROM

IF AUX OUT

TO IF B IN

C A U T I O N

The maximum power input to any of the input connectors (IF A IN, IF B IN, or 
REF A IN 100 MHz) must not 10 dBm!

Содержание Phase Matrix PXI-1440

Страница 1: ...Santa Clara CA 95054 Tel 1 408 610 6810 www phasematrix com PXI MICROWAVE IF DOWNCONVERTER MODULE Quick Start Guide Quick Start Guide PXI 1440 CCN 13001 Manual Assy Part Number 5585114 01 Manual Text...

Страница 2: ...to the United States National Institute of Standards and Tech nology Manual Change Information As Phase Matrix Inc continually improves and updates its products changes to the material covered by the...

Страница 3: ...s to describe the installation process of the stand alone hardware and software for the PXI 1440 and to confirm basic module operation This document describes information for the PXI 1440 when used in...

Страница 4: ...optional CD or from the Phase Matrix web site at www phasematrix com Screen Resolution at least 1024 768 pixels recommended Adobe Reader 6 0 1 or later to view or search PDF versions of manuals Interf...

Страница 5: ...be installed in any hybrid PXI slot marked with a peripheral slot compatibility image the letter H and a solid circle containing the slot number Do not install the PXI 1440 into a slot for PXI Express...

Страница 6: ...hassis Figure 1 Install PXI Modules into a PXI Chassis Phase Matrix PXI modules are sensitive instruments that should be handled carefully Do not expose modules to temperatures or humidity beyond thei...

Страница 7: ...the injector ejector handle is pushed down in the unlatched downward position c Slide the module completely into the chassis When you begin to feel resistance push up on the injector ejector handle to...

Страница 8: ...optic cable Laptop computer control of PXI consists of an ExpressCard PCMCIA CardBus card in the laptop and a PXI PXI Express module in slot 1 of the PXI system connected by a copper cable PXI Remote...

Страница 9: ...PXI consists of an ExpressCard PCMCIA CardBus card in the laptop and a PXI PXI Express module in slot 1 of the PXI system connected by a copper cable 1 If the PXI remote controller PC or laptop comput...

Страница 10: ...d controller or the PXI remote controller being used has previously had a device driver of the same type installed it must be removed before installing a new PXI 1440 s device driver Refer to Step 3 3...

Страница 11: ...32 bit exe or PXI 14XX Driver 64 bit exe The PXI 14XX Series Devcie Driver Installer will open with a welcome window similar to the one shown in Figure 2 Figure 2 Example Welcome Window 4 Select Next...

Страница 12: ...o not accept the agreement you will not be able to install the software A Windows security window similar to the one shown in Figure 4 will appear Figure 4 Example of Windows Security Window 6 Select...

Страница 13: ...to the setup exe file and select Open to start the installation wizard After a small delay a screen similar to Figure 6 is displayed The actual screen that appears will have the label of the module So...

Страница 14: ...ontinue to the next screen Figure 7 Example of Selecting Destination Directory for Executables 4 Enter the destination directory for the PXI 1440 s Soft Front Panel executables or browse to a destinat...

Страница 15: ...re License Agreement If you agree to the terms of the software license agreement select I accept the License Agreement and select Next to continue If you do not agree you will not be able to install t...

Страница 16: ...16 Phase Matrix PXI 1440 IF Downconverter Module Quick Start Guide Figure 10 Example of Completing Soft Front Panel Installation Wizard 8 Select Finish to complete the Soft Front Panel installation...

Страница 17: ...select Device Manger 3 Select the View pull down menu 4 Select Devices by connection and the device should appear under NI VISA PXI Devices with the label as PM14XX NOTE If this is the first time tha...

Страница 18: ...ight click on the selected device and a popup menu is displayed with five selectable choices 6 Select Uninstall and a popup window is displayed with a conformation to continue with the uninstall proce...

Страница 19: ...Phase Matrix PXI 1440 IF Downconverter Module 19 Quick Start Guide Step 4 Review Hardware and Software Front Panels...

Страница 20: ...nt Panel of the PXI 1440 IF Downconverter Module CAUTION The PXI 1440 IF Downconverter Module is designed to be used with the PXI 1420 MW Downconverter Module and PXI 1430 RF Downconverter Module The...

Страница 21: ...his Solid AMBER LED blinks to indicate each time the module is written to or read from IF OUT SMB m connector is used to deliver an IF output signal out of the front panel The IF output signal is supp...

Страница 22: ...ombined video output of both signals that can be input into IF A IN and IF B IN these IF input signals can be supplied by the PXI 1430 RF Downconverter Module and PXI 1420 MW Downconverter Module IF A...

Страница 23: ...nformation cannot be changed through the Soft Front Panel Stop Button This button is used to stop the Soft Front Panel from reading and writing to the PXI module it halts the program and is the prefer...

Страница 24: ...ected the IF input signal that is applied to either IF A IN or IF B IN is used to deliver a narrow band IF signal output centered on 21 4 MHz out of the front panel The input signal at IF A IN or IF B...

Страница 25: ...d is within tolerance 3Volts 3 V power supply displays green when functioning and is within tolerance 12Volts 12 V power supply displays green when functioning and is within tolerance LO3 Sense Indica...

Страница 26: ...b displays unit information such as model number serial number configuration control number FPGA part number EEPROM part number and last calibration date The bottom portion of this tab displays calibr...

Страница 27: ...ually improves and updates its products changes to materials occur When a part or assembly in a Phase Matrix instrument is changed to the extent that it is no longer interchangeable with an earlier pa...

Страница 28: ...orizontal Slider Bar to view correction values to the left and right of the values currently being displayed IF B IN to IF OUT Loss This table displays the correction data associated with the IF B IN...

Страница 29: ...response Use the Horizontal Slider Bar to view correction values to the left and right of the values currently being displayed 30 kHz Filter Response This table displays the correction data associated...

Страница 30: ...30 Phase Matrix PXI 1440 IF Downconverter Module Quick Start Guide This Page Left Intentionally Blank...

Страница 31: ...requency This process confirms that all connectors switches cables and related circuitry used in these signal paths are functional Equipment Requirements Figure 16 shows a block diagram of the equipme...

Страница 32: ...2 MHz RBW 1 MHz VBW Auto Attenuation 10 dB Scale 1 dB Div Reference 0 dB PXI 1450B Local Oscillator Module Frequency Initialized to default settings The PXI 1450B Local Oscillator Module is only used...

Страница 33: ...le to REF A IN of the PXI 1440 IF Downconverter Module Connect the IF AUX OUT of the PXI 1440 IF Downconverter Module to the RF Input of the Spectrum Analyzer 2 Set the Signal Generator frequency to 2...

Страница 34: ...N Set Select Output Path to IF OUT Wide BW 5 Connect the Spectrum Analyzer to IF OUT 6 Confirm that the output frequency on the Spectrum Analyzer is 249 4 MHz and the power level is 2 dBm 3 dB 7 Repea...

Страница 35: ...or s cavity and use a dry cotton swab to absorb remaining alcohol Never use abrasives to clean connectors Storing PXI units that have SMAs Place a protective cap on connectors when not in use Never se...

Страница 36: ...must be included in any correspondence regarding your PXI module A typical serial number label is shown in Figure 17 Figure 17 Example of a Label Affixed to the Side of the PXI Module Factory Service...

Страница 37: ...tainer cannot be used pack the PXI module in a heavy 275 pound test double walled carton with approximately four inches of packing material between the PXI module and the inner carton 4 Seal the carto...

Страница 38: ...r electrical equipment for measurement control and laboratory use EN 55011 2007 Industrial scienti c and medical ISM radio frequency Equipment Electromagnetic disturbance characteristics Limits and me...