Equipment Setup to Confirm Basic Module Operation
Figure 16
Equipment Setup to Confirm Basic Module Operation
IF Center Frequency and Bandwidth Check
1
Connect the equipment as shown in
•
Connect the RF Output of the Signal Generator
to RF B IN of the PXI-1430B.
•
Connect the LO OUT of the PXI-1450B Local Oscillator Module
to LO B IN of the PXI-1430B.
•
Connect the REF B OUT of the
PXI-1450B Local Oscillator Module
to REF B IN of the PXI-1430B.
•
Connect the IF B OUT of the PXI-1430B
to the RF Input of the Spectrum Analyzer.
2
Set the Signal Generator frequency to 1.5 GHz and the output power
to -40 dBm.
3
Set the LO frequency to 4.75 GHz.