42
|
ni.com
|
NI 5114 Calibration Procedure
Figure 1.
Complete Calibration Programming Flow
Appendix B: Calibration Utilities
NI-SCOPE supports several calibration utilities you can use to retrieve information about
adjustments performed on the NI 5114, change the external calibration password, and store
small amounts of information in the onboard EEPROM. Although you can retrieve some data
using MAX, you can retrieve all the data programmatically using NI-SCOPE functions.
MAX
To retrieve data using MAX, complete the following steps:
1.
Select the device from which you want to retrieve information from
My System»Devices
and Interfaces
.
2.
Select the
Settings
tab in the lower left corner.
You should see information about the last date and temperature for both external and
self-calibration.
Document
Post-Adjustment
Results
Document
Pre-Adjustment
Results
Calibration/
Verification
Complete
Yes
No
Self-Calibrate
Verify
Meets
Calibration
Test Limits?
Review Verification/
Adjustment Procedure
or Return Device
Adjust (Calibration Dates
and Temperatures Updated)
Verify