Figure 4.  

TB-4340 Circuit Board Parts Locator Diagram

Note

Refer to the 

NI PXIe-4340 User Manual

 for wiring diagrams showing how to 

connect sensors to the TB-4340.

6.

Tighten the terminal screws to a torque of 0.2 to 0.25 N · m (1.77 to 2.21 lb · in.).

Caution

Any wires connected to the ground must be sufficiently insulated from 

high voltage.

7.

Use the ground lugs to attach a shield wire to the ground.

Note

Refer to the 

NI PXIe-4340 User Manual

 for details about shielding the 

signals.

8.

Reinstall the strain-relief bar and tighten the strain-relief screws.

9.

Use tie wraps to connect the shielded signal wires to the tie-wrap holes for additional strain 
relief when necessary.

10. Reinstall the top cover and tighten the captive top cover screws.

Note

For information about sensors, go to 

ni.com/sensors

.

C1

B1

A1

COPYRIGHT 2015

S/N

159026B-01L

NI TB-4340

EX+

EX–

RS+

RS–

AI+

AI–

CT

AIGND

CHANNEL 0

EXITATION

VERIFICATION

EX+

EX–

RS+

RS–

AI+

AI–

CT

AIGND

CHANNEL 2

EXITATION

VERIFICATION

EX+

EX–

RS+

RS–

AI+

AI–

CT

AIGND

CHANNEL 1

EX+

EX–

RS+

RS–

AI+

AI–

CT

AIGND

CHANNEL 3

PFIO

DGND

R11

R10

R13

R12

R15

R14

R17

R16

C

Содержание NI TB-4340

Страница 1: ...s that could be damaged by the voltage present on the terminal block However you should only use these terminal blocks with their supported modules Contents Electromagnetic Compatibility Guidelines 2...

Страница 2: ...ions when the product is connected to a peripheral device or test object or if the product is used in residential or commercial areas To minimize interference with radio and television reception and p...

Страница 3: ...re Note You must be an administrator to install NI software and devices on your computer Ensure that the following software is installed before installing the PXIe 4340 and TB 4340 hardware 1 Your app...

Страница 4: ...d PXI Express slot in the chassis PXIe 4340 modules can be placed only in PXI Express Peripheral slots PXI Express Hybrid Peripheral slots and PXI Express System Timing slots Refer to the chassis docu...

Страница 5: ...ng cable shields To connect signals to the terminal block refer to Figures 3 and 4 while completing the following steps Note You can find the pinout names and locations in Measurement Automation Explo...

Страница 6: ...determine the function of the terminal The NI PXIe 4340 User Manual provides more detailed wiring information Make sure no exposed wire extends past the screw terminal Exposed wire increases the risk...

Страница 7: ...Note Refer to the NI PXIe 4340 User Manual for details about shielding the signals 8 Reinstall the strain relief bar and tighten the strain relief screws 9 Use tie wraps to connect the shielded signal...

Страница 8: ...f the terminal block 3 Tighten the mounting screw at the top of the TB 4340 to attach it to the PXIe 4340 module Note For safety purposes and to prevent damage to equipment when high voltages are pres...

Страница 9: ...figuring channels and tasks and getting started developing an application For detailed NI software version support refer to the NI DAQmx Readme 1 Launch MAX 2 Expand Devices and Interfaces and expand...

Страница 10: ...tions without installing hardware Refer to the Create a Simulated Device section for instructions for creating NI DAQmx simulated devices 3 Right click the module name and select Self Test When the se...

Страница 11: ...ni com support Take an NI DAQmx Measurement NI DAQmx Channels and Tasks Refer to the NI DAQmx Help for complete information about channels and tasks Use the DAQ Assistant to configure virtual channels...

Страница 12: ...Help Find Examples in your application software Text based code examples are located at All Programs National Instruments NI DAQmx Text Based Code Support Modify example code and save it in an applic...

Страница 13: ...release using a flathead screwdriver to disengage the latch b With the latch release raised grasp the terminal block and pull it away from the module Figure 7 Removing the TB 4340 from the PXIe 4340 M...

Страница 14: ...e captive screws on the top and bottom of the module Caution Do not pull the front panel to remove the module Doing so may cause module components to come into contact with adjacent modules causing da...

Страница 15: ...documentation by right clicking your module in MAX and selecting Help Online Device Documentation A browser window opens to ni com manuals with the results of a search for relevant documents If you do...

Страница 16: ...y dry and free from contaminants before returning it to service Environmental Maximum altitude 2 000 m 800 mbar at 25 C ambient temperature Pollution Degree 2 Indoor use only Operating Environment Amb...

Страница 17: ...the requirements of MIL PRF 28800F Class 3 Safety Measurement Category I1 Caution Do not connect the TB 4340 to signals or use for measurements within Measurement Categories II III or IV Caution The p...

Страница 18: ...is any industrial scientific or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection analysis purposes Note For the standards appli...

Страница 19: ...m troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extende...

Страница 20: ...s You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal expor...

Отзывы: