background image

Refer to the 

NI Trademarks and Logo Guidelines

 at 

ni.com/trademarks

 for information on National Instruments trademarks.

Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location: 

Help

»

Patents

 in your software, the

patents.txt

 file on your media, or the 

National Instruments Patent Notice

 at 

ni.com/patents

. You can find information about

end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the 

Export

Compliance Information

 at 

ni.com/legal/export-compliance

 for the National Instruments global trade compliance policy and

how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the
applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2015 National Instruments. All rights reserved.

375319A-01

May15

Содержание NI PXIe-5646R-G

Страница 1: ...mpaired if it is used in a manner not described in this document Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Verifying the Kit Contents...

Страница 2: ...Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance operate this product only with cables le...

Страница 3: ...ument Other Equipment There are several required items not included in your device kit that you need to install or operate the NI 5646R G Required Items A PXI Express chassis and chassis documentation...

Страница 4: ...ator to install NI software on your computer 1 Install an ADE such as LabVIEW or LabWindows CVI 2 Insert the driver software media into your computer The installer should open automatically If the ins...

Страница 5: ...chassis to rise above the optimal operating temperature for the device potentially causing thermal shutdown shorter lifespans or improper performance 5 Position the chassis so that inlet and outlet ve...

Страница 6: ...ting screws increases mechanical stability and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 13 Cover all empty slots u...

Страница 7: ...r switch routed to the NI 5646R G ensure no signal transients are sourced to the NI 5646R G Hardware Front Panel Connectors and Indicators Figure 4 NI 5646R G Front Panel DIGITAL I O LVTTL REF OUT 1 6...

Страница 8: ...erence Clock REF OUT Output terminal that can export a 10 MHz Reference Clock or the 250 MHz Sample Clock PFI 0 Programmable function digital I O DIO connector for use with triggers or events DIGITAL...

Страница 9: ...more information about front panel connectors and LEDs Configuring the NI 5646R G in MAX Use Measurement Automation Explorer MAX to configure your National Instruments hardware MAX informs other prog...

Страница 10: ...temperature has drifted more than 5 C from the temperature at the last self calibration To periodically adjust for small performance drifts that occur with product aging NI recommends you perform the...

Страница 11: ...cation in Microsoft Visual C C Related Information For detailed instructions about how to generate signals in a specific ADE refer to the Getting Started section of the NI RF Signal Generators Help Re...

Страница 12: ...ED on the NI 5646R G hardware module illuminates 5 Click RF On Off to stop signal generation Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical su...

Страница 13: ...sing a PXI controller verify that a National Instruments entry appears in the system device list Reinstall NI RFSG and the device if error conditions appear in the list If you are using an MXI control...

Страница 14: ...software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located using the NI Example Finder Located at ni...

Страница 15: ...your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressw...

Страница 16: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Отзывы: