Caution
Do not disconnect the cable that connects CAL IN to CAL OUT.
Removing the cable from or tampering with the CAL IN or CAL OUT front
panel connectors voids the product calibration and specifications are no longer
warranted.
Note
Do not install a device if it appears damaged in any way.
3. Unpack any other items and documentation from the kit.
Store the device in the antistatic package when the device is not in use.
Preparing the Environment
Ensure that the environment you are using the NI 5644R in meets the following specifications.
Operating ambient temperature
(IEC 60068-2-1, IEC 60068-2-2)
0 °C to 55 °C
Operating relative humidity
(IEC 60068-2-56)
10% to 90%, noncondensing
Maximum altitude
2,000 m (800 mbar) (at 25 °C ambient
temperature)
Pollution Degree
2
Indoor use only.
Caution
Clean the hardware with a soft, nonmetallic brush or lint free cloth. Make
sure that the hardware is completely dry and free from contaminants before
returning it to service.
Note
Refer to the
NI PXIe-5644R Specifications
at
specifications.
NI PXIe-5644R Getting Started Guide
|
© National Instruments
|
3