background image

If the installation window does not appear, navigate to the drive, double-click it, and

double-click 

autorun.exe

.

6. Follow the instructions in the installation prompts.

Note

 

Windows users may see access and security messages during

installation. Accept the prompts to complete the installation.

Note

 

The default installation installs NI LabVIEW Instrument Design

Libraries for Vector Signal Transceivers, NI-RFmx, NI-RFSA, and NI-RFSG.

7. When the installer completes, select 

Restart

 in the dialog box that prompts you to restart,

shut down, or restart later.

Installing the NI 5644R

Caution

 

To prevent damage to the NI 5644R caused by ESD or contamination,

handle the module using the edges or the metal bracket.

You must install the software before installing the hardware.

Before you install the hardware, refer to the guidelines in the 

Maintain Forced-Air Cooling

Note to Users

 included with the module to ensure that the device can cool itself effectively.

This document is also available at 

ni.com/manuals

.

The NI 5644R is a three-slot module with two backplane connectors. The module must be
installed into three adjacent chassis slots, and the left two slots must be PXI Express
compatible.
1. Ensure the AC power source is connected to the chassis before installing the module.

The AC power cord grounds the chassis and protects it from electrical damage while you

install the module.

2. Power off the chassis.
3. Inspect the slot pins on the chassis backplane for any bends or damage prior to

installation. Do not install a module if the backplane is damaged.

4. If the chassis has multiple fan speed settings, ensure the fans are set to the highest setting.

Note

 

Inadequate air circulation could cause the temperature inside the chassis

to rise above the optimal operating temperature for the device, potentially

causing thermal shutdown, shorter lifespans, or improper performance.

5. Position the chassis so that inlet and outlet vents are not obstructed.

Caution

 

Do not disconnect the cable that connects CAL IN to CAL OUT.

Removing the cable from or tampering with the CAL IN or CAL OUT front

panel connectors voids the product calibration and specifications are no longer

warranted.

6. Remove the black plastic covers from all the captive screws on the module front panel.
7. Identify a supported slot in the chassis. The following figure shows the symbols that

indicate the slot types.

8

  | 

ni.com

  | 

NI PXIe-5644R Getting Started Guide

Содержание NI PXIe-5644R

Страница 1: ...PXIe 5644...

Страница 2: ...this equipment may be impaired if it is used in a manner not described in the documentation Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2...

Страница 3: ...ocumentation Furthermore any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Caution To...

Страница 4: ...that the environment you are using the NI 5644R in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56...

Страница 5: ...at you need to operate the NI 5644R Your application may require additional items not included in your kit to install or operate your device Required Items A PXI Express chassis and chassis documentat...

Страница 6: ...re Software Options NI provides three software options for programming the NI 5644R NI RFmx NI RFSA and NI RFSG instrument driver software and the instrument design libraries NI PXIe 5644R Getting Sta...

Страница 7: ...data Use NI RFSG to generate continuous or modulated waveforms Use NI RFSA and NI RFSG with the NI Modulation Toolkit to develop software defined radio transmitters and receivers Use NI RFSG with the...

Страница 8: ...I RFSA and NI RFSG Soft Front Panels support Yes Yes Yes LabVIEW examples or sample projects Includes LabVIEW examples Includes LabVIEW examples Includes LabVIEW sample projects with additional exampl...

Страница 9: ...jacent chassis slots and the left two slots must be PXI Express compatible 1 Ensure the AC power source is connected to the chassis before installing the module The AC power cord grounds the chassis a...

Страница 10: ...e the base engages with the guides in the chassis Figure 3 NI 5644R Module Installation 1 2 3 ESD SENSIT IVE NI PX Ie 5 644 R Vec tor Sig nal Tran sce iver 65 MH z 6 0 GHz 1 PXI Express Chassis 2 NI P...

Страница 11: ...evice is powered off may cause damage Ensure you are properly grounded when manipulating cables or antennas connected to the NI 5644R RF IN connector If you are using nonisolated devices such as a non...

Страница 12: ...5 MHz 6 0 GHz REF IN 5Vp p MAX 50 RF IN 0 RF OUT 0 CAL IN CAL OUT LO OUT LO OUT LO IN LO IN RF IN RF OUT REF OUT 1 65Vp p 50 DIGITAL I O LVTTL ESD SENSITIVE 33 dBm MAX 50 0 dBm 50 0 3 dBm 50 Reverse P...

Страница 13: ...z Reference Clock REF OUT Output terminal that can export a 10 MHz Reference Clock or the 120 MHz Sample Clock PFI 0 Programmable function digital I O DIO connector for use with triggers or events DIG...

Страница 14: ...le from or tampering with the CAL IN or CAL OUT front panel connectors voids the product calibration and specifications are no longer warranted Connector that is used when running self calibration on...

Страница 15: ...een The device is generating a waveform Dim amber The device is waiting for an acquisition Reference Trigger Solid amber The device is acquiring a waveform Solid red The device has detected an error T...

Страница 16: ...erform a complete self calibration after first installing your module and letting it warm up for 30 minutes Note Warm up begins when the PXI Express chassis has been powered on and the operating syste...

Страница 17: ...tory and import library files located in one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit NIDir Shared ExternalCompilerSupport C include Windows 10 64 bit 8 1 64 bit 7 64 bit 32...

Страница 18: ...s x86 IVI Foundation IVI Microsoft Visual C C Use the header files located in the IVIROOTDIR32 Include directory and import library files located in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 i...

Страница 19: ...ocated in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 is one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit Program Files IVI Foundation IVI Windows 10 64 bit 8 1 64 bit 7 64...

Страница 20: ...ming the NI 5644R 1 Launch LabVIEW 2 In the New Project dialog box select Sample Projects in the left pane 3 Navigate to and open the project that best matches your application requirements Using Micr...

Страница 21: ...continuous wave CW tone RF signal at the RF output terminal of your device 5 Select Start All Programs National Instruments NI RFSA NI RFSA Soft Front Panel to launch the NI RFSA Soft Front Panel 6 S...

Страница 22: ...isconnect any signals from the PXI Express module front panels Caution Do not disconnect the cable that connects CAL IN to CAL OUT Removing the cable from or tampering with the CAL IN or CAL OUT front...

Страница 23: ...controller verify that a National Instruments entry appears in the system device list If error conditions appear in the list right click the NI PXIe 5644R and NI RIO FPGA Device nodes and select Updat...

Страница 24: ...Support ni com support more about your products through ni com Located online at ni com manuals Located online at ni com gettingstarted RF Solutions ni com rf NI PXIe 5644R Specifications NI RF Vecto...

Страница 25: ...formation support phone numbers email addresses and current events Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names...

Отзывы: