22. Commit the settings to hardware using the niRFSA Initiate VI or the
niRFSA_Initiate
function.
23. Wait 200 ms.
24. Measure the
Channel A Power
. Calculate
RF Power
using the following equation:
RF Power
=
Channel A Power
+
RF Input Calibration Factor
Use the appropriate RF frequency for the channel A sensor. Calculation of the
RF Input
Calibration Factor
may require linear interpolation.
25. Measure and record the channel B power. This value is the
IF Power
. Make this
measurement using the appropriate IF frequency for the channel B sensor.
26. Calculate
Gain
using the following formula:
Gain
=
IF Power
-
RF Power
27. Store the appropriate calibration constants using the niRFSA Cal Adjust IF Attenuation
Calibration VI or the
niRFSA_CalAdjustIFAttenuationCalibration
function.
You must provide the following information:
•
IF filter
•
Attenuation settings in a two-element array [IF1, IF2]
•
Gain
from step 26
28. Repeat steps 21 to 27 for all possible combinations of IF attenuation settings from the valid
values in Table 22.
Note
NI recommends keeping your RF signal generator enabled when repeating
steps unless making a connection change. Keeping the signal generator enabled
reduces the wear on the RF signal generator relays. After making a connection
change and enabling your generator, wait 5 seconds for the system to settle.
29. Repeat steps 14 to 28 for all
IF Filters
in Table 21.
30. Close the NI 5601 IF attenuation calibration step using the niRFSA Close Calibration
Step VI or the
niRFSA_CloseCalibrationStep
function.
31. Close the external calibration by calling the niRFSA Close Ext Cal VI with the
action
parameter set to
Commit
or by calling the
niRFSA_CloseExtCal
function with the
action
parameter set to
NIRFSA_VAL_EXT_CAL_COMMIT
.
32. Close the niRFSG session using the niRFSG Close VI or the
niRFSG_close
function.
Table 22.
IF Attenuator Settings
IF1 Attenuator
IF2 Attenuator
0 dB to 15 dB in 1 dB steps
0 dB to 15 dB in 1 dB steps