NI PXIe-5601 Calibration Procedure
|
© National Instruments
|
13
11. If you used the spectrum analyzer, set the reference level, frequency span, and resolution
bandwidth according to Table 4. Set the center frequency according to Table 5.
12. Set the RF source to the same frequency you set for the NI 5601 center frequency according
to Table 3.
13. At each NI 5601 center frequency and reference level setting, perform the following steps:
a.
Adjust the RF source power, in dBm, using the following equation:
Channel A Power
+
RF Input Attenuated Calibration Factor
=
Reference Level
±1 dB
When making the adjustment, use the appropriate RF frequency for the channel A
sensor. Calculation of the
RF Input Attenuated Calibration Factor
may require linear
interpolation.
Note
Set the signal generator to its maximum power if it does not have sufficient
power to reach the reference level.
b.
Record the channel A power.
c.
Measure the channel B power or spectrum analyzer power using one of the following
methods:
•
If you are using the power meter, measure and record the channel B power. Use
the appropriate IF frequency from Table 5 for the channel B sensor.
•
If you are using the spectrum analyzer, measure and record the
Spectrum
Analyzer Power
. Correct the
Spectrum Analyzer Power
using the appropriate
IF Calibration Factor
for the IF frequency in Table 5 and the following formula:
Corrected Spectrum Analyzer Power
=
Spectrum Analyzer Power
+
IF Calibration Factor
Table 4.
Spectrum Analyzer Settings
Parameter
Value
Reference Level
0 dBm
Frequency Span
10 Hz
Resolution Bandwidth
200 Hz
Table 5.
Spectrum Analyzer Center Frequency Settings
NI 5601 Center Frequency
IF Frequency (Spectrum Analyzer
Center Frequency)
10 MHz to <120 MHz
187.5 MHz
120 MHz to <330 MHz
53 MHz
330 MHz to 6.6 GHz
187.5 MHz