Figure 5.
Current Remote Sense Output Verification Connection Diagram
Sense LO
Sense HI
HI
LO
Load2
Load1
DMM
Current
Mode
+
–
Lead Drop 1
Lead Drop 2
NI-DCPower Device
2.
Set the
niDCPower Sense
property or
NIDCPOWER_ATTR_SENSE
attribute to Remote.
3.
Set the
niDCPower Output Function
property or
NIDCPOWER_OUTPUT_FUNCTION
attribute to DC Current for the NI 4141.
Verifying Current Remote Sense Accuracy
Use the NI 4141 in constant current mode with a test circuit to simulate the voltage drop
between the device and a load.
Measure all of the specified resistors using the DMM before you connect equipment. Record
the values.
Refer to the following table as you complete the following steps.
Table 5.
Current Remote Sense Output Verification
Level Range
Limit Range and Limit
Test Point
Load
1
Load
2
Base Limit
10 µA
10 V
-10 µA
100 kΩ
100 kΩ
3.3 nA
0 µA
10 µA
100 μA
10 V
-100 μA
10 kΩ
10 kΩ
33 nA
0 μA
100 μA
1 mA
10 V
-1 mA
1 kΩ
1 kΩ
330 nA
0 mA
1 mA
NI PXIe-4141 Calibration Procedure
|
© National Instruments
|