Table 5. Current Offset Verification
Level
Range
Limit Range
Limit
Test
Point
As-Found Offset
Test Limit
As-Left Offset Test
Limit
600 mV
1 μA
1 μA
0 mV
±40 pA
±17 pA
10 µA
10 µA
±300 pA
±130 pA
100 µA
100 µA
±2 nA
±970 pA
1 mA
1 mA
±20 nA
±9.7 nA
10 mA
10 mA
±200 nA
±97 nA
100 mA
100 mA
±2 μA
±970 nA
1 A
1 A
±20 μA
±8.5 µA
3 A
1 A
±600 μA
±85 µA
1.
Disconnect all equipment from the output of the PXIe-4137.
2.
Measure the internal device temperature and perform self-calibration if necessary.
a)
If the internal device temperature exceeds T
cal
±1 °C, wait up to five minutes for the
temperature to stabilize to within T
cal
±1 °C.
b)
If after five minutes the stable temperature still exceeds T
cal
±1 °C, call the self-
calibration VI or function.
3.
Take a current measurement using the PXIe-4137.
4.
Record the value from the previous step.
5.
Verify that the recorded value falls within the test limits.
6.
If more than one limit range is specified, repeat the previous steps using the values
specified in each limit range.
Verifying Load Regulation
Note
Although load regulation is listed as a typical specification for the
PXIe-4137, verification is required. If the PXIe-4137 fails the load regulation
verification procedure, discontinue use of the device and contact an authorized NI
service representative to request a Return Material Authorization (RMA).
Refer to the following table as you complete the following steps:
Table 6. Load Regulation Verification
Level Range
Limit Range and Limit
Test Point
As-found/As-left Limit
10 mA
600 mV
10 mA
2 mV
1.
Set the
niDCPower Output Function
property or
NIDCPOWER_OUTPUT_FUNCTION
attribute to DC Current for the PXIe-4137.
12
|
ni.com
|
NI PXIe-4137 Calibration Procedure