Adjusting AC Voltage
Note
You must adjust DC voltage before you adjust AC voltage.
Note
If you do not use the AC voltage modes for any measurements, or the
accuracy of these modes is irrelevant, you can skip this section in the calibration
procedure and go directly to the
Adjusting Current
section.
Complete the following steps to adjust the AC voltage of the device:
1.
Reset the calibrator.
2.
Connect the PXIe-4080/4082 to the calibrator as shown in
on page 6.
3.
Call the niDMM Cal Adjust Misc VI or the
niDMM_CalAdjustMisc
function with
Type
set to
NIDMM_EXTCAL_MISCCAL_VREF_AC_VOLTAGE
.
4.
Adjust the AC Voltage gain with each configuration in the following table:
a)
Commit the device to specified range.
b)
Configure the calibrator as specified and allow it to settle.
c)
Call the niDMM Cal Adjust AC Filter VI or the
niDMM_CalAdjustACFilter
function with the specified parameters.
DMM Settings
Calibrator Settings
Function
Range
Amplitude
Frequency
AC Volts
50 mV
50 mV
1 kHz
5 kHz
20 kHz
50 kHz
100 kHz
200 kHz
300 kHz
500 kHz
PXIe-4080/4082 Calibration Procedure
|
© National Instruments
|
23