National Instruments NI PXI-665 Series Скачать руководство пользователя страница 7

©

 National Instruments Corporation

7

NI PXI-665x Calibration Procedure

c.

Call niSync Connect Clock Terminals VI to connect the oscillator 
to 

ClkOut

.

d.

Call niSync Close VI to close the session handle.

3.

Complete steps a–c to measure the oscillator frequency through PFI 0. 
To obtain an accurate measurement for calibration, the measurement 
duration must be made sufficiently long. Refer to Table 1 for the 
measurement duration to use.

LabVIEW Block Diagram

NI-SYNC C Function Call

Call 

niSync_ConnectClkTerminals

 

with the following parameters:

vi

“<

SessionHandle

>”

sourceTerminal

NISYNC_VAL_OSCILLATOR

destinationTerminal

NISYNC_VAL_CLKOUT

LabVIEW Block Diagram

NI-SYNC C Function Call

Call 

niSync_close

 with the following 

parameter:

vi

“<

SessionHandle

>”

Table 1.  

Measurement Duration

Device

Measure Accuracy

Required Measurement 

Duration

NI PXI-6652

200 ppb

0.5 seconds

NI PXI-6653

1 ppb

100 seconds

Содержание NI PXI-665 Series

Страница 1: ...g the Clock Source 4 Measuring the Frequency 6 Adjustment 10 Reverification 13 Where to Go for Support 14 Conventions The following conventions apply to this document The symbol leads you through nest...

Страница 2: ...ns Software This calibration procedure requires NI Sync and NI VISA NI Sync and NI VISA support a number of application development environments ADEs and programming languages including LabVIEW LabWin...

Страница 3: ...the NI PXI 665x in Slot 2 of the PXI chassis The calibration procedure requires features of PXI that are accessible only in slot 2 Use of an NI PXI 1044 or NI PXI 1045 chassis is recommended to reduc...

Страница 4: ...e driver and application software is installed on the host computer 5 Configure the hardware properly with Measurement Automation Explorer MAX Refer to the NI PXI 665x Installation Guide for details a...

Страница 5: ...tion Call Call niSync_init with the following parameters resourceName Dev1 idQuery NULL resetDevice VI_TRUE vi SessionHandle LabVIEW Block Diagram NI SYNC C Function Call Call niSync_SetAttributeViBoo...

Страница 6: ...nc Initialize VI to set up a handle for the device b Set a writable NI Sync property node to pass TRUE to the ClkOut Gain Enable attribute LabVIEW Block Diagram NI SYNC C Function Call Call niSync_clo...

Страница 7: ...on must be made sufficiently long Refer to Table 1 for the measurement duration to use LabVIEW Block Diagram NI SYNC C Function Call Call niSync_ConnectClkTerminals with the following parameters vi Se...

Страница 8: ...agram NI SYNC C Function Call Call niSync_init with the following parameters resourceName Dev1 idQuery NULL resetDevice VI_FALSE vi SessionHandle LabVIEW Block Diagram NI SYNC C Function Call Call niS...

Страница 9: ...r period following a calibration Table 2 One Year Test Limits Device Specified Accuracy Aging Acceptable Frequency Range As Found Low Limit High Limit NI PXI 6652 TCXO 3 5 ppm 9 999 965 Hz 10 000 035...

Страница 10: ...plete steps a c to set the oscillator control voltage The range of acceptable voltage values is 0 0 V to 4 095 V with frequency increasing as voltage increases Use a control voltage of 2 0475 V which...

Страница 11: ...quency section To obtain an accurate measurement for calibration the measurement duration must be made sufficiently long Refer to Table 1 Measurement Duration for the measurement duration to use Compl...

Страница 12: ...Ranges 5 Commit the calibration values to the Calibration EEPROM using the following procedure a Call niSync Initialize External Calibration VI to initialize the process Note NI is the default user p...

Страница 13: ...on section Re verifying after making the adjustments ensures that the NI PXI 665x is operating within its test limits LabVIEW Block Diagram NI SYNC C Function Call Call niSync_InitExtCal with the foll...

Страница 14: ...s National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your sup...

Отзывы: