NI 5922 Calibration Procedure

9.

Calculate the error in timing as parts per million (ppm) using the following formula:

where

a

 = the measured frequency

Compare the result to the 

Calibration Test Limits

 or the 

Published Specifications

 listed in 

Table 7. If the result is within the selected test limit, the device has passed this portion of 
the verification.

Adjusting the NI 5922

If the NI 5922 successfully passed each of the verification procedures within the calibration test 
limits, then an adjustment is recommended but not required to guarantee its published 
specifications for the next two years. If the oscilloscope was not within the calibration test limits 
for each of the verification procedures, you can perform the adjustment procedure to improve 
the accuracy of the oscilloscope. Refer to the 

External Calibration Options

 section to determine 

which procedures to perform.

An adjustment is required only once every two years. Following the adjustment procedure 
automatically updates the calibration date and temperature in the EEPROM of the oscilloscope.

Note

If the oscilloscope passed the entire verification procedure within the 

calibration test limits and you do not want to perform an adjustment, you can update 
the calibration date and onboard calibration temperature without making any 
adjustments by completing 

only

 steps 4 and 13 in this section.

Complete all of the following steps to externally adjust the NI 5922.
1.

Connect the scope calibrator to the channel 0 input of the oscilloscope. Configure the 
calibrator to output an exact 1 MHz sine wave with 1 V peak-to-peak amplitude and 1 M

 

load impedance.

2.

Connect the DC source calibrator to the channel 1 input of the oscilloscope. Configure the 
DC source calibrator to output an exact 4 V DC signal into a 1 M

 load impedance 

(channel 1).

3.

Wait 3,000 ms for the calibrator to settle.

4.

Call 

niScope_CalStart

 (niScope Cal Start VI) using the user password. The factory 

default password for the NI 5922 is 

NI

.

5.

Call 

niScope_CalSelfCalibrate

 (niScope Cal Self Calibrate VI) with the following 

parameters:

channelList

VI_NULL

option

VI_NULL

Table 7.  

NI 5922 Timing Accuracy

Calibration Test Limit

Published Specification

±5.3 ppm

 ±50 ppm

error

a

2,000,000

2

-------------------------------------

=

Содержание NI PXI-5922

Страница 1: ...libration 2 External Calibration 2 Calibration Intervals 2 Software and Documentation 2 Documentation 4 Self Calibration Procedures 4 MAX 5 Scope SFP 5 NI SCOPE 5 External Calibration Options 6 Comple...

Страница 2: ...nsate for environmental changes Caution Although you can use self calibration repeatedly self calibrating the NI 5922 more than a few times a day may cause excessive wear on the relays over time The m...

Страница 3: ...cription IVI Bin niscope_32 dll NI SCOPE driver containing the entire NI SCOPE API including calibration functions IVI Lib msc niscope lib NI SCOPE library for Microsoft C containing the entire NI SCO...

Страница 4: ...Self calibrate the oscilloscope before you perform verification NI SCOPE includes self calibration example programs for LabVIEW CVI and Microsoft Visual C You can initiate self calibration using Meas...

Страница 5: ...o the oscilloscope 2 Launch the Scope SFP by selecting Start Programs National Instruments NI SCOPE Scope Soft Front Panel 3 Select the device you want to calibrate using the Select Device dialog box...

Страница 6: ...to limits given in Tables 3 through 7 This document provides two sets of test limits for most verification stages the calibration test limits and the published specifications The calibration test lim...

Страница 7: ...ication fall within the calibration test limits the device is guaranteed to meet or exceed its published specifications for a full calibration interval In this case you can update the calibration date...

Страница 8: ...gure 2 Optional Calibration Programming Flow Self Calibrate Verify Yes No Meets Published Specifications Yes Results Within Published Specs Adjustment Optional No Adjust Cal Dates and Temperatures Upd...

Страница 9: ...extra noise that can affect measurements Keep relative humidity between 10 and 90 non condensing Maintain an ambient temperature of 23 5 C Allow a warm up time of at least 15 minutes after NI SCOPE l...

Страница 10: ...is section begin with niScope_init niScope Initialize VI with resetDevice set to NISCOPE_VAL_TRUE and end with niScope_close niScope Close VI If any of these tests fail immediately after you perform a...

Страница 11: ...hat corresponds to the vertical range used If the result is within the selected test limit the device has passed this portion of the verification 7 Repeat steps 1 through 6 for each iteration in Table...

Страница 12: ...r vi The instrument handle you obtained from niScope_init 6 Call niScope_FetchArrayMeasurement niScope Fetch Array Measurement VI with the following parameters channelList 0 timeout 5 0 arrayMeasFunct...

Страница 13: ...st 0 inputImpedance NISCOPE_VAL_50_OHM maxInputFrequency 0 3 Call niScope_ConfigureVertical niScope Configure Vertical VI with the following parameters channelList 0 range 2 offset 0 0 coupling NISCOP...

Страница 14: ...the following parameters channelList 0 inputImpedance NISCOPE_VAL_1_MEG_OHM maxInputFrequency 0 8 Call niScope_InitiateAcquisition niScope Initiate Acquisition VI with the following parameter vi The...

Страница 15: ...t 0 0 coupling NISCOPE_VAL_DC probeAttenuation 1 0 enabled NISCOPE_VAL_TRUE 3 Call niScope_ConfigureHorizontalTiming niScope Configure Horizontal Timing VI with the following parameters minSampleRate...

Страница 16: ...a percentage of input using the following formula where a the measured positive voltage b the measured negative voltage c the applied positive voltage d the applied negative voltage Compare the resul...

Страница 17: ...ing niScope Configure Horizontal Timing VI with the following parameters minSampleRate 15 000 000 minNumPts 1 000 000 refPosition 50 0 numRecords 1 enforceRealtime NISCOPE_VAL_TRUE 4 Call niScope_Comm...

Страница 18: ...ion date and temperature in the EEPROM of the oscilloscope Note If the oscilloscope passed the entire verification procedure within the calibration test limits and you do not want to perform an adjust...

Страница 19: ...le 10 Call niScope_CalAdjustRange niScope Cal Adjust Range VI with the following parameters channelName 1 range 0 stimulus 4 0 11 Disconnect or disable all inputs to the oscilloscope 12 Call niScope_C...

Страница 20: ...nformation from My System Devices and Interfaces 2 Select the Calibration tab You should see information about the last date and temperature for both external and self calibration NI SCOPE NI SCOPE pr...

Страница 21: ...is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from...

Отзывы: