28
|
ni.com
|
NI PXI-5404 Calibration Procedure
Closing the Calibration Session
If a function returns an error, call
niHSSources_CalEnd
(niHSSources Cal CalEnd VI) to
abort and close the session. Set the following parameters:
•
sessionHandle
—The handle of the calibration session for the module
•
action
—
niHSSources_ABORT
If you do not receive an error, call
niHSSources_CalEnd
(niHSSources Cal CalEnd VI) to
save the new constants. Set the following parameters:
•
sessionHandle
—The handle of the calibration session for the module
•
action
—
niHSSources_COMMIT_CONSTANTS
Calibration Function Reference
This section provides detailed descriptions of NI 5404 calibration and constant functions that
support signal source performance. The NI HSSources API provides access to the
niarbcal.dll
and can be used to write calibration procedures for any NI 5404 module.
Note
All functions return signed, 32-bit integers corresponding to the status of the
operation.