NI PXI-5404 Calibration Procedure
|
© National Instruments
|
3
Table 1.
Equipment Required for Calibrating NI 5404 Modules
Instrument
Applicable
Specification
Minimum
Specifications
Recommended
Instrument(s)
Digital multimeter
(DMM)
SINE out amplitude
accuracy
AC amplitude accuracy
better than ±0.1% at
50 kHz at 4 V
pk-pk
NI PXI-4070
Agilent (HP) 34401A
Keithley 2000
Male banana to female
BNC adapter
—
—
Male BNC to female
SMB cable
50
Ω
—
Spectrum analyzer/
Frequency Counter
Frequency accuracy
Bandwidth >150 MHz
Frequency accuracy to
±500 ppb
NI PXI-5660
Agilent/HP 8560E
Agilent/HP 53131A
or HP 53132A
with timebase option
001, 010, or 012
Rohde & Schwarz
FSU26
Male SMA to female
SMB cable
—
—
Oscilloscope
CLOCK out duty
cycle accuracy
Bandwidth
≥
500 MHz
Timebase Accuracy:
≤
200 ppm
Tektronix TDS3054
Male BNC to female
SMB cable
—
—
Power meter and thermal
power sensor
SINE out amplitude
passband flatness
±0.12 dB accuracy for
flatness measurements
from 9 kHz to 105 MHz
Rohde & Schwarz
NRVS meter
Rohde & Schwarz
NRV-Z51 or
Rohde & Schwarz
NRV-Z53 sensor
Rohde & Schwarz
NRP-Z91
Female N-Type to
female SMB cable
50
Ω
—