National Instruments Corporation
33
NI PXI-4070 Calibration Procedure
34. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Resolution
=
0.1
35. Call
niDMM_Read
. Subtract the previously stored 100 k
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
36. Output 10 k
Ω
on the calibrator with 2-wire compensation turned on,
but with external sense turned off.
37. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e3
•
Resolution
=
0.01
38. Call
niDMM_Read
. Subtract the previously stored 10 k
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
39. Output 1 k
Ω
on the calibrator with 2-wire compensation turned on,
but with external sense turned off.
40. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e3
•
Resolution
=
1e–3
41. Call
niDMM_Read
. Subtract the previously stored 1 k
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
42. Output 100
Ω
on the calibrator with 2-wire compensation turned on,
but with external sense turned off.
43. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100
•
Resolution
=
100e–6
44. Call
niDMM_Read
. Subtract the previously calculated 100
Ω
range
offset from this measurement. Verify that the result falls between the
tolerances listed in Table 14.