Caution

Never

 connect the AC power cable to the power supply shuttle until you install it 

in a PXI-1006 chassis. Do 

not

 use, test, or configure the power supply shuttle outside of a 

chassis.

The power supply shuttle is a replacement part for the PXI-1006 chassis. 
The 

NI PXI-1006 User Manual

 contains other chassis service procedures. 

The chassis includes a hardcopy of the 

NI PXI-1006 User Manual

Additionally, you can download a softcopy from 

ni.com/support

.

Removal

Before attempting to replace the power supply shuttle, verify that there is 
adequate clearance behind the chassis. Set the power switch on the front 
panel to the Standby position and disconnect the power cable from the 
power supply shuttle on the back of the chassis. Identify the ten mounting 
screws that attach the power supply shuttle to the chassis. Refer to Figure 
for the mounting screw locations. Using a Phillips screwdriver, loosen the 
captive screws. Pull on the two rear handles of the power supply shuttle to 
remove it from the back of the chassis.

Installation

Ensure that there is no visible damage to the new power supply shuttle. 
Verify that the housing and connector on the new power supply shuttle 
have no foreign material inside. Install the new power supply shuttle into 
the opening on the rear of the chassis. Tighten the ten mounting screws with 
a Phillips screwdriver.

Configuration

The fan-speed selector switch is on the rear panel of the power supply 
shuttle. Refer to Figure to locate the fan-speed selector switch. Select HI 
for maximum cooling performance (recommended) or AUTO for quieter 
operation.