NI 5922 Calibration Procedure
Documentation
Consult the following documents for information about the NI 5922 and your application
software. All documents are available on
and help files install with the software.
Self-Calibration Procedures
The NI 5922 includes precise internal circuits and references used during self-calibration to
adjust for time and temperature drift. The oscilloscope gain, offset, and flatness are all corrected
in self-calibration. Measuring the accuracy of these internal calibration circuits with another
instrument during external calibration provides traceability for the verification procedure.
Absolute accuracy is ensured by compensating for any error measured in the internal references.
Note
Self-calibrate the oscilloscope before you perform verification. NI-SCOPE
includes self-calibration example programs for LabVIEW, CVI, and
Microsoft Visual C.
You can initiate self-calibration using Measurement & Automation Explorer (MAX), using the
Scope Soft Front Panel (SFP), or programmatically using NI-SCOPE.
NI PXI/PCI-5922 Getting Started Guide
Contains: NI-SCOPE installation, hardware installation, and hardware
programing instructions
NI PXI/PCI-5922 Specifications
Contains: NI 5922 specifications and calibration intervals
NI-SCOPE Readme
Contains: Operating system and application software support in NI-SCOPE
NI High Speed Digitizers Help
Contains: Detailed information about NI PXI/PCI-5922
NI-SCOPE LabVIEW Reference Help
Contains: LabVIEW programming concepts and reference information about
NI-SCOPE VIs and functions
NI-SCOPE Function Reference Help
Contains: Reference information for NI-SCOPE C functions and NI-SCOPE
C properties