NI 5105 Calibration Procedure
9.
Fetch a waveform from the device and perform a voltage average measurement using the
niScope Fetch Measurement Poly VI. Select the Measurement Scalar DBL instance of
the VI.
Compare the resulting average voltage to the value listed in the
Published Specifications
column in Table 5 that corresponds to the settings used.
10. Repeat steps 2 through 9 for each iteration listed in Table 5.
11. Move the calibrator test head to the next device input channel.
12. Repeat steps 2 through 11 for each of the remaining channels, updating the
channelList
parameter with the appropriate channel number.
13. End the session using the niScope Close VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_Fetch
Measurement
with the
following parameters:
timeout
:
1.0
vi
: The instrument handle from
niScope_init
channelList
: The name of the
channel being tested
scalarMeasFunction
:
NISCOPE_VAL_VOLTAGE
_AVERAGE
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the
following parameters:
vi
: The instrument handle from
niScope_init