National Instruments Corporation
13
AO Waveform Calibration Procedure for NI-DAQmx
4.
Perform an external calibration adjustment using
DAQmxESeriesCalAdjust
(
DAQmx Adjust AO-Series
Calibration VI
).
5.
Save the adjustment to the EEPROM, or onboard memory, using
DAQmxCloseExtCal
(
DAQmx Close External Calibration
). This
function also saves the date, time, and temperature of the adjustment
to the onboard memory.
6.
Disconnect the calibrator from the device.
The device is now calibrated with respect to your external source.
After adjusting the device, you may want to verify the analog output
operation. To do this, repeat the steps in the
section using the 24-hour test limits in the
AO Device Test Limits
section.
NI-DAQ Function Call
LabVIEW Block Diagram
Call
DAQmxAOSeriesCalAdjust
with the following parameters:
calHandle
:
calHandle
referenceVoltage
:
5
NI-DAQ Function Call
LabVIEW Block Diagram
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_
Action_Commit