Channel-to-channel

None

Channel-to-earth ground

Continuous

60 V DC, Measurement Category I

Withstand

1,000 V RMS, verified by a 5 s dielectric withstand test

Caution

 Do not connect the product to signals or use for measurements

within Measurement Categories II, III, or IV.

Attention

 Ne pas connecter le produit à des signaux dans les catégories

de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans

ces catégories.

Warning

 Do not connect the product to signals or use for measurements

within Measurement Categories II, III, or IV, or for measurements on MAINs

circuits or on circuits derived from Overvoltage Category II, III, or IV which

may have transient overvoltages above what the product can withstand.

The product must not be connected to circuits that have a maximum

voltage above the continuous working voltage, relative to earth or to

other channels, or this could damage and defeat the insulation. The

product can only withstand transients up to the transient overvoltage

rating without breakdown or damage to the insulation. An analysis of

the working voltages, loop impedances, temporary overvoltages, and

transient overvoltages in the system must be conducted prior to making

measurements.

Mise en garde

 Ne pas connecter le produit à des signaux dans les

catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans

ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de

surtensions de catégorie II, III ou IV pouvant présenter des surtensions

transitoires supérieures à ce que le produit peut supporter. Le produit

ne doit pas être raccordé à des circuits ayant une tension maximale

supérieure à la tension de fonctionnement continu, par rapport à la

ni.com

4

NI-9477 Getting Started

Содержание NI-9477

Страница 1: ...NI 9477 Getting Started 2022 07 06...

Страница 2: ...dous Locations 5 Electromagnetic Compatibility Guidelines 7 Preparing the Environment 7 NI 9477 Pinout 8 Connecting Digital Devices 8 Increasing Current Drive 9 Protecting the Digital Outputs from Fly...

Страница 3: ...ety and EMC ratings for the entire system Safety Guidelines Caution Observe all instructions and cautions in the user documentation Using the product in a manner not specified can damage the product a...

Страница 4: ...uits that have a maximum voltage above the continuous working voltage relative to earth or to other channels or this could damage and defeat the insulation The product can only withstand transients up...

Страница 5: ...parts of equipment circuits powered by regulated low voltage sources and electronics Note Measurement Categories CAT I and CAT O are equivalent These test and measurement circuits are for other circu...

Страница 6: ...Each NI 9477 is marked II 3G and is suitable for use in Zone 2 hazardous locations in ambient temperatures of 40 C Ta 70 C If you are using the NI 9477 in Gas Group IIC hazardous locations you must us...

Страница 7: ...ception and prevent unacceptable performance degradation install and use this product in strict accordance with the instructions in the product documentation Furthermore any changes or modifications t...

Страница 8: ...he device datasheet on ni com manuals for complete specifications NI 9477 Pinout DO16 DO17 DO18 DO19 DO20 DO21 DO22 DO23 COM COM DO24 DO25 DO26 DO27 DO28 DO29 DO30 DO31 DO0 DO1 DO2 DO3 DO4 DO5 DO6 DO7...

Страница 9: ...ered on Ensure that the devices you connect to the NI 9477 are compatible with the output specifications of the NI 9477 Refer to the device datasheet at ni com manuals for output specifications Increa...

Страница 10: ...Power Supply Flyback Diode Where to Go Next NI Services Visit ni com support to find support resources including documentation downloads and troubleshooting and application development self help such...

Отзывы: