![National Instruments NI 9269 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/national-instruments/ni-9269/ni-9269_getting-started-manual_709258005.webp)
Safety Voltages
Channel-to-channel
Continuous
250 Vrms,
Measurement Category II
Withstand
1,390 Vrms, verified by a 5 s
dielectric withstand test
Channel-to-earth ground
Continuous
250 Vrms,
Measurement Category II
Withstand
2,300 Vrms, verified by a 5 s
dielectric withstand test
Division 2/Zone 2
hazardous locations
applications (Channel-to-
channel and channel-to-
earth ground)
60 VDC,
Measurement Category I
Measurement Category I is for measurements performed on
circuits not directly connected to the electrical distribution system
referred to as
MAINS
voltage. MAINS is a hazardous live
4
|
ni.com
|
NI 9269 Getting Started Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com