National Instruments Corp.
25
NI 9237 Operating Instructions and Specifications
Rejection at oversample rate
1
f
s
= 10 kS/s.................................. 60 dB @ 640 kHz
f
s
= 50 kS/s.................................. 90 dB @ 3.2 MHz
Common-mode voltage,
all signals to earth gro
u
nd................. ±60 VDC
CMRR
Relative to earth gro
u
nd
2
(
f
in
= 0 to 60 Hz) ......................... 140 dB
Relative to EX–
(
f
in
= 0 to 1 kHz) ......................... 85 dB
SFDR (1 kHz, –60 dBFS)................. 106 dB
Total Harmonic Distortion (THD)
1 kHz, –20 dBFS ........................ 100 dB
8 kHz, –20 dBFS ........................ 90 dB
1
Rejection by analog prefilter of signal freq
u
encies at oversample rate.
2
Meas
u
red with a balanced cable. Shielded cables that are not twisted-pair may be
significantly
u
nbalanced. To improve the balance of shielded, twisted-pair cables,
NI recommends twisting together the AI+/AI
–
pair, the RS+/RS
–
pair, and the
EX+/EX
–
pair.