![National Instruments NI 9237 Скачать руководство пользователя страница 16](http://html2.mh-extra.com/html/national-instruments/ni-9237/ni-9237_getting-started-manual_3545732016.webp)
Connecting TEDS Sensors
You can connect TEDs sensors to the NI 9237.
Figure 5.
Connecting TEDS Sensors to the NI 9237
T+
Data
COM
T–
NI 9237
TEDS
Ensure that neither the TEDS data (T+) nor the TEDS return (T-)
signal is tied in common to any AI signals on the NI 9237. The
NI 9237 connects all the T- signals together internally. The
NI 9237 with DSUB has only three T- pins. To connect four
TEDS sensors to the NI 9237 with DSUB, wire the TEDS return
signals of two of the sensors to one of the T- pins.
NI 9237 Getting Started Guide
|
© National Instruments
|
15