CALIBRATION PROCEDURE

NI 9229/9239

4-Channel, ±60 V/±10 V, 24-Bit Simultaneous, Channel-to-Channel 
Isolated Analog Input Modules

This document contains the verification and adjustment procedures for the National 
Instruments 9229/9239. For more information about calibration solutions, visit 

ni.com/

calibration

.

Contents

Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4

Accuracy Verification....................................................................................................... 4
Gain Matching Verification.............................................................................................. 6
Phase Matching Verification ............................................................................................ 7

Adjustment................................................................................................................................ 9

Accuracy Adjustment ....................................................................................................... 9

EEPROM Update ..................................................................................................................... 10
Reverification ........................................................................................................................... 10
Accuracy Under Calibration Conditions .................................................................................. 10
Worldwide Support and Services ............................................................................................. 11

Software

Calibrating the NI 9229/9239 requires the installation of NI-DAQmx 9.5 or later on the 
calibration system. You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx 

supports LabVIEW, LabWindows

/CVI

, ANSI C, Microsoft Visual C++, Microsoft Visual 

Basic .NET, and Microsoft Visual C#. When you install NI-DAQmx you only need to install 
support for the application software that you intend to use.

Содержание NI 9229

Страница 1: ...y Verification 4 Gain Matching Verification 6 Phase Matching Verification 7 Adjustment 9 Accuracy Adjustment 9 EEPROM Update 10 Reverification 10 Accuracy Under Calibration Conditions 10 Worldwide Sup...

Страница 2: ...nstructions and Specifications NI 9229 9239 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabV...

Страница 3: ...ctions to the device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minut...

Страница 4: ...elf Test to ensure that the device is working properly Verification The following performance verification procedures describe the sequence of operation and provide test points required to verify the...

Страница 5: ...Accuracy Verification Module Physical Channels Min V Max V Scaled Units Terminal Configuration 9229 AI0 3 60 60 Volts Differential 9239 AI0 3 10 10 Volts Differential Table 3 NI 9229 9239 Timing Conf...

Страница 6: ...ure function generator according to Table 6 3 Acquire and average samples a Create and configure an AI voltage task on the NI 9229 9239 according to Table 7 Table 6 Function Generator Configuration fo...

Страница 7: ...If the Gain Matching verification procedure determines that the NI 9229 9239 is outside of the limits refer to Worldwide Support and Services for assistance in returning the device to NI Phase Matchin...

Страница 8: ...able the function generator g Clear the task 5 Calculate the phase for each channel on the NI 9229 9239 NI recommends using the Extract Single Tone Information VI to calculate phase Table 9 Function G...

Страница 9: ...r as shown in Figure 1 Note If the calibrator outputs are truly floating the negative output must be connected to a quiet earth ground as well as AI to give the entire system a ground reference 3 Adju...

Страница 10: ...determine the As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to...

Страница 11: ...om manuals Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and appl...