NI 9216/9226 Calibration Procedure

6. Connect the power source to the cDAQ-9178 chassis.
7. Launch Measurement & Automation Explorer (MAX).
8. Right-click the device name and select 

Self-Test

 to ensure that the module is working 

properly.

Verification

The following performance verification procedure describes the sequence of operation and test 

points required to verify the NI 9216/9226. The verification procedure assumes that adequate 

traceable uncertainties are available for the calibration references.

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9216/9226.
1. Set the calibrator to Standby mode (STBY).
2. Connect the desired channel of the NI 9216/9226 to the calibrator as shown in Figure 1. 

Connect only one channel at a time.

Figure 1.  

4-Wire Resistance Connections

3. Set the calibrator as a resistance source to a Test Point value indicated in Table 2.
4. Set calibrator COMP to 4-wire.

Table 2.  

NI 9216/9226 Accuracy Verification Test Limits

Module

Range

Test Point

1-Year Limits

Minimum 

(

Ω

)

Maximum 

(

Ω

)

Value (

Ω

)

Upper 

Limit (

Ω

)

Lower 

Limit (

Ω

)

NI 9216

0

400

0

0.03

-0.03

100

100.042

99.958

320

320.0684

319.9316

NI 9226

0

4000

0

0.3

-0.3

1000

1000.43

999.57

3200

3200.72

3199.28

EX+

HI

LO

AUX LO

AUX HI

EX–

RTD–

RTD+

NI 9216/9226

CALIBRATOR

4-Wire Connection

Содержание NI 9226

Страница 1: ...NI 9226...

Страница 2: ...est Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 7 Reverification 7 Accuracy Under Calibration Conditions 7 Wo...

Страница 3: ...NI 9216 circuitry and specifications NI 9226 Getting Started Guide NI 9226 connection information NI 9226 Datasheet NI 9226 circuitry and specifications NI DAQmx Readme Operating system and applicati...

Страница 4: ...midity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9216 9226 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set...

Страница 5: ...llowing procedure to determine the As Found status of the NI 9216 9226 1 Set the calibrator to Standby mode STBY 2 Connect the desired channel of the NI 9216 9226 to the calibrator as shown in Figure...

Страница 6: ...x in AIx For example use AI3 for channel 3 c Configure the task timing according to Table 4 d Start the task e Read the samples from the NI 9216 9226 f Average the samples g Stop and clear the task 7...

Страница 7: ...ain an array of recommended calibration resistance values d Set the calibrator to a reference value determined by the array of recommended calibration voltages e Set calibrator COMP to 4 wire f Set th...

Страница 8: ...djustment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Accuracy Under Calibratio...

Страница 9: ...ined in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 Wo...

Отзывы: