NI-9222 Connection Guidelines

 Make sure that devices you connect to the NI-9222 are compatible with the

module specifications.

 You must use 2-wire ferrules to create a secure connection when connecting

more than one wire to a single terminal on the NI-9222 with screw terminal.

Wiring for High-Vibration Applications

If your application is subject to high vibration, NI recommends that you follow these

guidelines to protect connections to the NI-9222 with screw terminal:

Use ferrules to terminate wires to the detachable connector.

Use the NI-9971 backshell kit.

Cable Requirements for EMC Compliance

Select and install cables for the NI-9222 with screw terminal in accordance with the

following requirements:

Connect the cable shield to the chassis ground (grounding screw of the

chassis) using the shortest length of wire possible.

Use shielded, twisted-pair cables (Belden 9451 or equivalent).