Note
The TB-9212 contains a plastic insert to prevent
accidental wire contact with the metal enclosure.
Safety Voltages
Connect only voltages that are within the following limits:
Channel-to-channel isolation
Up to 2,000 m altitude
Continuous
250 Vrms, Measurement
Category II
Withstand
1,500 Vrms, verified by a 5 s
dielectric test
Up to 5,000 m altitude
Continuous
60 VDC, Measurement
Category I
Withstand
1,000 Vrms, verified by a 5 s
dielectric test
4
|
ni.com
|
NI 9212 Getting Started Guide