NI 9209 Calibration Procedure
|
© National Instruments
|
9
Accuracy Under Calibration Conditions
The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.
The following accuracy table is valid for calibration under the following conditions:
•
Ambient temperature 23 °C ± 5 °C
•
NI 9209 installed in slot 8 of an NI cDAQ-9178 chassis
•
Slots 1 through 7 of the NI cDAQ-9178 chassis are empty
Note
For operational specifications, refer to the most recent
NI 9209 Datasheet
online at
ni.com/manuals
.
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At
you have access to everything from troubleshooting and application development
self-help resources to email and phone assistance from NI Application Engineers.
for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit
to register your National Instruments product. Product registration
facilitates technical support and ensures that you receive important information updates from NI.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For
telephone support in the United States, create your service request at
or dial
1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the
Worldwide Offices section of
to access the branch office websites, which
provide up-to-date contact information, support phone numbers, email addresses, and current
events.
Table 8.
NI 9209 Accuracy Under Calibration Conditions
Input Channels
Percentage of Reading
Percentage of Range
*
Differential
0.097%
0.0055%
Single-Ended
0.093%
0.0053%
*
Range equals 10.4 V for voltage.