NI 9209 Calibration Procedure

|

© National Instruments

|

9

Accuracy Under Calibration Conditions

The values in the following table are based on calibrated scaling coefficients, which are stored 
in the onboard EEPROM.

The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

NI 9209 installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Note

For operational specifications, refer to the most recent 

NI 9209 Datasheet

 

online at 

ni.com/manuals

.

Worldwide Support and Services

The National Instruments website is your complete resource for technical support. At 

ni.com/

support

 you have access to everything from troubleshooting and application development 

self-help resources to email and phone assistance from NI Application Engineers.

Visit 

ni.com/services

 for NI Factory Installation Services, repairs, extended warranty, and 

other services.

Visit 

ni.com/register

 to register your National Instruments product. Product registration 

facilitates technical support and ensures that you receive important information updates from NI. 

National Instruments corporate headquarters is located at 11500 North Mopac Expressway, 
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For 
telephone support in the United States, create your service request at 

ni.com/support

 or dial 

1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the 
Worldwide Offices section of 

ni.com/niglobal

 to access the branch office websites, which 

provide up-to-date contact information, support phone numbers, email addresses, and current 
events.

Table 8.  

NI 9209 Accuracy Under Calibration Conditions

Input Channels

Percentage of Reading

Percentage of Range

*

Differential

0.097%

0.0055%

Single-Ended

0.093%

0.0053%

Range equals 10.4 V for voltage.

Содержание NI 9209

Страница 1: ...rential Voltage Accuracy Verification 4 Single Ended Voltage Accuracy Verification 6 Adjustments 7 Voltage Accuracy Adjustment 7 EEPROM Update 8 Reverification 8 Accuracy Under Calibration Conditions...

Страница 2: ...NI 9209 specific information NI 9209 Datasheet NI 9209 specifications and calibration interval NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW Help LabVIEW...

Страница 3: ...iminate noise and thermal offsets Maintain an ambient temperature of 23 5 C The device temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of...

Страница 4: ...librator as shown in Figure 1 Figure 1 Differential Voltage Accuracy Connections Note If the calibrator outputs are truly floating connect the negative output to a quiet earth ground as well as COM to...

Страница 5: ...Set the calibrator to Standby mode STBY 9 Repeat steps 3 through 8 for each test point listed in Table 4 10 Disconnect the calibrator from the device 11 Repeat steps 1 through 10 for each voltage chan...

Страница 6: ...e to a Test Point value indicated in Table 6 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage task on the NI 9209 according to Table 5 b Conf...

Страница 7: ...red to adjust the NI 9209 Voltage Accuracy Adjustment Complete the following procedure to adjust the voltage accuracy performance of the NI 9209 1 Set the calibrator to Standby mode STBY 2 Connect the...

Страница 8: ...de the physical channel s should be a union of the channels specified in both configurations n Close and commit the calibration session 4 Remove the connections from the NI 9209 EEPROM Update When an...

Страница 9: ...rs Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your National Instruments product Product registration faci...

Страница 10: ...nts You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal exp...

Отзывы: