NI 9205 Operating Instructions and Specifications
22
ni.com
AI Absolute Accuracy Tables and Formulas
The val
u
es in the following tables are based on calibrated scaling
coefficients, which are stored in the onboard EEPROM.
Acc
u
racy s
u
mmary
Nominal
Range (V)
Absolute Accuracy
at Full Scale
*
(
μ
V)
Random Noise,
σ
(
μ
Vrms)
Sensitivity
†
(
μ
V)
±10
6,230
240
96.0
±5
3,230
116
46.4
±1
690
26
10.4
±0.2
174
10
4.0
*
Absol
u
te acc
u
racy val
u
es at f
u
ll scale on the analog inp
u
t channels ass
u
me the
device is operating within 70 °C of the last external calibration and are valid for
averaging 100 samples immediately following internal calibration. Refer to the
for more information.
†
Sensitivity is the smallest voltage change that can be detected. It is a f
u
nction of
noise.