NI 9203 Calibration Procedure
Accuracy Under Calibration Conditions
The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.
The following accuracy table is valid for calibration under the following conditions:
•
Ambient temperature 23 °C ±5 °C
•
NI 9203 installed in slot 8 of an NI cDAQ-9178 chassis
•
Slots 1 through 7 of the NI cDAQ-9178 chassis are empty
Note
For operational specifications, refer to the most recent
NI 9203 Operating
Instructions/User Guide and Specifications
online at
ni.com/manuals
.
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At
you have access to everything from troubleshooting and application development
self-help resources to email and phone assistance from NI Application Engineers.
for NI Factory Installation Services, repairs, extended warranty, and
other services.
to register your National Instruments product. Product registration
facilitates technical support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European
Communities using the manufacturer’s declaration of conformity. This system affords the user
protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
. If your product supports calibration,
you can obtain the calibration certificate for your product at
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For
telephone support in the United States, create your service request at
Table 8.
NI 9203 Accuracy Under Calibration Conditions
Input Range
Percent of Reading
(Gain Error)
Percent of Range*
(Offset Error)
Unipolar
0.080% max
0.019%
†
max
Bipolar
0.082% max
0.033% max
*
Range equals 21.5 mA for Unipolar and 43 mA (±21.5 mA) for Bipolar.
†
This specification is lower than the Percent of Range typical specification in the
NI 9203
Operating Instructions and Specifications
because the typical specification shows digits only
to the hundredth place.