background image

NI 6711/6713/6731/6733 Calibration Procedure

8

ni.com

6.

Call

Select_Signal

with

signal

and

source

set to

ND_GPCTR0_OUTPUT

to route the counter signal to the GPCTR0_OUT

pin on the device I/O connector.

7.

Call

GPCTR_Control

with

action

set to

ND_PROGRAM

to start the

generation of the square wave. The device begins to generate a 5 MHz
square wave when

GPCTR_Control

completes execution.

8.

Compare the value read by the counter to the test limits shown in the
appropriate table in the

Specifications

section. If the value falls

between these limits, the device has passed this test.

9.

Disconnect the counter from the device.

You have now verified the device counter.

Adjusting the NI 6711/6713/6731/6733

This procedure adjusts the AO calibration constants. At the end of each
calibration procedure, these new constants are stored in the factory area
of the device EEPROM. An end-user cannot modify these values, which
provides a level of security that ensures users do not accidentally access or
modify any calibration constants adjusted by the metrology laboratory.

This step in the calibration process calls functions in NI-DAQ and in
the

ni671x.dll

. For further information on the functions in the

ni671x.dll

, refer to the comments in the

ni671x.h

file.

1.

Disconnect all cables to the device. Make sure the device is not
connected to any circuits other than those specified by the calibration
procedure.

2.

To internally calibrate the device, call the

Calibrate_E_Series 

function with the following parameters set as indicated:

calOP

set to

ND_SELF_CALIBRATE

setOfCalConst

set to

ND_USER_EEPROM_AREA

calRefVolts

set to

0

3.

Connect the calibrator to the device according to Table 2.

Table 2.

Connecting the Calibrator to the Device

6711/6713/6731/6733 Pins

Calibrator

EXTREF (pin 20)

Output High

AOGND (pin 54)

Output Low

Pin numbers are given for 68-pin connectors only. If you are using a 50-pin connector,
refer to the device documentation for signal connection locations.

Содержание NI 6731

Страница 1: ...ed for calibrating NI 6711 6713 6731 6733 devices Note Refer to ni com support calibrat mancal htm for a copy of the ni671xCal dll file What Is Calibration Calibration consists of verifying the measur...

Страница 2: ...en calibrated with respect to an external standard However external variables such as temperature can still affect measurements The new calibration constants are defined with respect to the calibratio...

Страница 3: ...imize connections and test conditions during calibration Keep connections to the NI 6711 6713 6731 6733 short Long cables and wires act as antennae picking up extra noise which can affect measurements...

Страница 4: ...w to configure the device in NI DAQ Refer to the NI 671X 673X User Manual for detailed installation instructions You can install this manual when you install NI DAQ 1 Power down the computer 2 Install...

Страница 5: ...DAQ User Manual for PC Compatibles at ni com manuals These two documents provide detailed information about using NI DAQ The function reference help includes information on the functions in NI DAQ Th...

Страница 6: ...those specified by the calibration procedure 2 To internally calibrate the device call the Calibrate_E_Series function with the following parameters set as indicated calOP set to ND_SELF_CALIBRATE set...

Страница 7: ...ifying the Performance of the Counter This procedure verifies the performance of the counter The NI 6711 6713 6731 6733 devices have only one timebase to verify so you only need to verify counter 0 Be...

Страница 8: ...nnot modify these values which provides a level of security that ensures users do not accidentally access or modify any calibration constants adjusted by the metrology laboratory This step in the cali...

Страница 9: ...om the device The device is now adjusted with respect to the external source After the device is adjusted you can verify the AO operation by repeating the Verifying Analog Output section Specification...

Страница 10: ...r the test point value should fall between the upper and lower limit values These limits are expressed in volts Counters Because you cannot adjust the resolution of the counter timers these values do...

Страница 11: ...Programs National Instruments NI DAQ NI DAQ Help and the NI DAQ User Manual for PC Compatibles at ni com manuals for additional information on the software structure Verifying Analog Output Figure 1 V...

Страница 12: ...CTR_Control deviceNumber 0 ND_RESET GPCTR_Set_Application deviceNumber 0 ND_PULSE_TRAIN_GNR GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_1 2 GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_2 2 Select...

Страница 13: ...djusting the NI 6711 6713 6731 6733 Figure 3 Adjusting the NI 6711 6713 6731 6733 Calibrate_E_Series deviceNumber ND_SELF_CALIBRATE ND_USER_EEPROM_AREA 0 Get_Cal_Date deviceNumber CalDate Calibrate_E_...

Отзывы: