![National Instruments NI 6115 Скачать руководство пользователя страница 93](http://html.mh-extra.com/html/national-instruments/ni-6115/ni-6115_user-manual_709719093.webp)
Appendix A
Specifications
©
National Instruments Corporation
A-13
Voltage Output
Ranges .................................................... ±10 V
Output coupling...................................... DC
Output impedance .................................. 50
Ω
±5%
Current drive .......................................... ±5 mA
Output stability....................................... Any passive load
Protection .............................................. Short-circuit to ground
Power-on output voltage
(before software loads calibration values)
NI 6115 ........................................... ±400 mV
NI 6120 ........................................... ±80 mV
Initial power-up glitch
Magnitude ....................................... ±2 V
Duration .......................................... 200 ms
Table A-5.
NI 6115 Analog Output DC Accuracy Information
Nominal
Range at
Full Scale
(V)
Absolute Accuracy
Relative Accuracy
% of Reading
Offset
(mV)
Temp
Drift
(%/°C)
Absolute
Acc. at Full
Scale (mV)
Theoretical Resolution
(mV)
24 Hrs
90 Days
1 Year
±10
0.0437
0.0445
0.0454
8.9
0.0006
13.5
4.88
Table A-6.
NI 6120 Analog Output DC Accuracy Information
Nominal
Range at
Full Scale
(V)
Absolute Accuracy
Relative Accuracy
% of Reading
Offset
(mV)
Temp
Drift
(%/°C)
Absolute
Acc. at Full
Scale (mV)
Theoretical Resolution
(µV)
24 Hrs
90 Days
1 Year
±10
0.0460
0.0468
0.0477
1.9
0.0006
6.7
305.2