National Instruments Corporation
21
NI 5922 Calibration Procedure
2.
Call
niScope_ConfigureVertical
(niScope Configure
Vertical VI) with the following parameters:
•
channelList
:
"0"
•
range
: The
Range
value for the current iteration in Table 8
•
offset
:
0.0
•
coupling
:
NISCOPE_VAL_DC
•
probeAttenuation
:
1.0
•
enabled
:
NISCOPE_VAL_TRUE
3.
Call
niScope_ConfigureHorizontalTiming
(niScope Configure
Horizontal Timing VI) with the following parameters:
•
minSampleRate
: The
Sample Rate
value for the current iteration
in Table 8
•
minNumPts
:
500
•
refPosition
:
50.0
•
numRecords
:
400
•
enforceRealtime
:
NISCOPE_VAL_TRUE
4.
Call
niScope_ConfigureTriggerEdge
(niScope Configure
Trigger Edge VI) with the following parameters:
•
triggerSource
:
"0"
•
level
:
–0.1
•
slope
:
NISCOPE_VAL_POSITIVE
•
triggerCoupling
:
NISCOPE_VAL_DC
•
holdoff
:
0
•
delay
:
0
5.
Call
niScope_Commit
(niScope Commit VI) with the following
parameter:
•
vi
: The instrument handle you obtained from
niScope_init
6.
Connect the scope calibrator to the channel 0 input of the digitizer.
Configure the calibrator as follows:
a.
Set the calibrator to rising edge (500 ps) output mode.
b.
Set the frequency to 500 Hz.
c.
Set the peak-to-peak voltage amplitude to one tenth of the vertical
range used in step 2.
d.
Set the load impedance of the calibrator to 50
Ω
.
7.
Wait 1,000 ms for the impedance matching of the calibrator to settle.