Appendix A
Specifications
©
National Instruments Corporation
A-5
Distortion
Timebase System
Reference clock...................................... 10 MHz
Clock accuracy (as master) .................... 10 MHz ±50 ppm
1 MS/s
400 kHz
–160 dBfs/
–104 dBfs
500 kS/s
200 kHz
–160 dBfs/
–107 dBfs
200 kS/s
80 kHz
–160 dBfs/
–111 dBfs
100 kS/s
40 kHz
–160 dBfs/
–114 dBfs
50 kS/s
20 kHz
–160 dBfs/
–117 dBfs
20 kS/s
8 kHz
–160 dBfs/
–121 dBfs
10 kS/s
4 kHz
–160 dBfs/
–124 dBfs
* 1<n<2
32
in oscilloscope mode
Sampling Frequency
SFDR for input
0 dBfs
SFDR for input
–20 dBfs
SFDR for input
–60 dBfs (typical)
100 MS/s
50 dB
50 dB
N/A
12.5 MS/s
65 dB
85 dB
125 dB
5 MS/s
70 dB
90 dB
130 dB
2.5 MS/s
75 dB
95 dB
135 dB
1 MS/s
85 dB
105 dB
145 dB
500 kS/s
90 dB
110 dB
150 dB
200 kS/s
100 dB
110 dB
160 dB
100 kS/s
100 dB
110 dB
160 dB
50 kS/s
100 dB
110 dB
160 dB
20 kS/s
100 dB
110 dB
160 dB
10 kS/s
100 dB
110 dB
160 dB
Sampling Frequency
Bandwidth
Noise Density
Total Noise
Hz
Hz
Hz
Hz
Hz
Hz
Hz