NI 5421 Specifications
26
ni.com
Markers
Table 9.
Specification
Value
Comments
Destinations
1. PFI<0..1> (SMB front panel connectors)
2. PFI<4..5> (DIGITAL DATA & CONTROL front panel
connector)
3.
NI PXI-5421
—PXI_Trig<0..6> (backplane connector)
NI PCI-5421
—RTSI<0..7>
—
Quantity
One Marker per Segment.
—
Quantum
Marker position must be placed at an integer multiple of
four samples.
—
Width
>150 ns. Refer to t
m2
at
NI Signal Generators Help»
Devices»NI 5421»NI <bus>-5421»Waveform
Generation»Marker Events
.
—
Skew
Destination
With Respect to
Analog Output
With Respect to
Digital Data
Output
Refer to t
m1
at
NI Signal
Generators
Help»Devices»
NI 5421»
NI <bus>-5421»
Waveform
Generation»
Marker Events
.
PFI<0..1>
±2 Sample Clock
Periods
N/A
PFI<4..5>
N/A
<2 ns
NI PXI-5421
PXI_Trig<0..6>
NI PCI-5421
RTSI<0..6>
±2 Sample Clock
Periods
N/A
Jitter
20 ps rms
—